8.
Remove the defective tube:
• Press the flaps backwards to release the tube.
• Pull the tube gently out of the holders and then sideways out of
the scanner.
9.
If you are replacing the transparencies tube, you need to disassemble
the tube from the sockets as well:
• Remove the screw (G) to release the left socket (H).
• Remove the right socket (J) by simply pulling it off (you may need
to use a small screwdriver to loosen this socket).
• Assemble the new tube with the sockets.
10. Insert the replacement tube and mount it into the holders making
sure that it is properly held in place by the flaps.
11. Connect the plug (F).
12. IMPORTANT: The "T" on the lamp board cover (K) indicates the
socket for the transparencies tube.
13. Lower the feed table (E).
14. Mount the side panels (C) and (D), the screws (B) and the front
panel (A) following the reverse of the procedure used when
removing them.
NOTE! When mounting the front panel (A) make sure to give it a
firm push at the bottom to click it properly into position.
15. Turn on the scanner and check function of the new tube.
User's Guide - Flextight 949
Hardware Reference - Flextight 949
The Light Tubes
27