Table 5 Relationship Between Temperature Of Ion Selective Electrode And Theoretical Slope; Table 6 Relationship Between Temperature Of Chloride Ion Selective Electrode; Impact Of Temperature - Horiba Scientific 6560S-10C Instruction Manual

Combination type chloride ion selective electrode
Table of Contents

Advertisement

Impact of temperature

The potential gradient measured by the ion electrode method is impacted
by the temperature of the sample solution in accordance with the Nernst
equation. The higher the temperature is, the larger the potential gradient is.
The lower the temperature is, the smaller the potential gradient is.
The standard solution in calibration and the sample solution in
measurement shall have almost the same temperature. Match the
temperature as close as possible.
Theoretical
slope
The optimal way to minimize the impact of temperature on the
measurement of the Cl
the temperature of the standard solution in calibration and the temperature
of the sample solution in measurement to ±1°C.
The following table shows the change in the potential gradient of the
chloride ion selective electrode depending on the temperature.
temperature of chloride ion selective electrode and theoretical slope
Temperature [°C] Theoretical slope of chloride ion selective electrode
0
10
20
25
30
40
50
64
Table 5 Relationship between temperature of ion
selective electrode and theoretical slope
Unit: mV/decade (mV/concentration change of 10 times)
0
Monovalent
54.20 56.18 58.16 59.16 60.15 62.13 64.11
ions
Divalent
27.10 28.09 29.08 29.58 30.07 31.07 32.06
ions
-
concentration is to reduce the difference between
Table 6 Relationship between
Temperature (°C)
10
20
25
[mV/decade]
-54.20
-56.18
-58.16
-59.16
-60.15
-62.13
-64.11
30
40
50

Advertisement

Table of Contents
loading

Table of Contents