Filling With Internal Solution - Horiba Scientific 6560S-10C Instruction Manual

Combination type chloride ion selective electrode
Table of Contents

Advertisement

F
illing with internal solution
Be sure to fill the electrode with internal solution as
described in the following procedure before use.
 Be sure to fill with internal solution before
use.
If measurement is performed in the state in
which there is not much internal solution,
an inflow of sample solution into the
electrode will result in accurate
measurements no longer being possible.
 Use the internal solution included in the
Note
Note
package or available as a consumable sold
separately by HORIBA Advanced Techno.
(
 Discharge the internal solution each time
measurement finishes.
Items to be prepared
Internal solution
1
Remove the protective cap.
(Except when it is already removed,
such as first time of use.)
12
"Consumables" (page 72))
Dropper
View video
View video
Protective cap