Horiba Scientific 6560S-10C Instruction Manual page 51

Combination type chloride ion selective electrode
Table of Contents

Advertisement

2
Remove the internal solution in one of the following ways.
<Discharging from the liquid junction>
Spread soft and clean lint-free paper or cloth
beneath the electrode and then press the
electrode against it from above.
The inside of the electrode moves up and
down.
Continue pressing from above until all of the
internal solution is discharged.
When pressing the electrode from above, do not press too
strong or have an impact on the electrode.
It may damage the electrode.
Note
Note
<Extracting from the internal solution fill port>
Turn the electrode upward and extract the
internal solution with a dropper.
When disposing of the internal solution and paper or cloth with
internal solution adhered to it, follow the local rules and
regulations.
Tip
Tip
The internal solution
oozes out from the
liquid junction.