Calibration Operation - Horiba Scientific 6560S-10C Instruction Manual

Combination type chloride ion selective electrode
Table of Contents

Advertisement

C
alibration operation
Measure standard solutions in ascending order of ion concentration to create
calibration curves in the meter (
electrode" (page 58)).
Perform calibration once every day or before measurement as appropriate.
Before operation
 Be sure to fill with internal solution and rinse the electrode
before the calibration operation. (
solution" (page 12), "Rinsing the electrode" (page 15))
 Perform conditioning before the first use, before use after
Note
Note
storage for two weeks or longer, or if the electrode is less
responsive.
(
Note on page 23)
Items to be prepared
Standard
solutions
 Prepare at least two types of standard
solution with different concentrations.
 Always use new standard solution for
calibration.
 After the standard solutions (50 mL) and
ISA (50 mL) included in the package are
Note
Note
used up, use the standard solution (500
mL) and ISA (500 mL) available as
consumables.
"Measurement principle of ion selective
ISA
Syringe
(50 mL)
"Filling with internal
19