Specifications - T&D TR-81 Instruction Manual

Thermo recorder
Hide thumbs Also See for TR-81:
Table of Contents

Advertisement

A:
Sometimes communication will not work if an extension cable has been added to the
communication cable or if a switch has been added to the serial port (RS-232C).
A:
Check to see if some other communication software is in use.
A:
Most desktop models have two serial ports, try changing the plug to the other port
and change the settings.
Q : : : : :
The images are hanging off the display screen. What should I do ?
A:
The display should be set at a resolution of 640x480 dpi or higher.
Q : : : : :
The date and the time of the recorded data are different from the actual date and
time. What should I do?
A:
The Thermo Recorder takes the date and time from your computer when it is set
up. Make sure that the date and time are correct on your computer.
Q : : : : :
Is the data backed up automatically?
A:
If the battery power goes low while recording the BAT lamp will come on. If the
power decreases even more, recording will be ceased and a backup of the re-
corded data will be made.

Specifications

Thermo Recorder TR-81
Recording Channels
2 Channels
Sensors
Pt100 Three Wire Sensors
Measurement Range
-200℃ to 600℃
± 0.3℃(‑200 〜 80℃)  ± 0.5℃(80 〜 450℃)
Measurement Reliability
(when main unit is at 20℃± 10℃)
± 1.0℃(450 〜 600℃)
※ No account for sensor differences related to sensor type.
Measurement Resolution
0.1℃
Measurement Current
1 mA
Starting Methods
Immediate / Programmed (by computer only)
Recording Modes
One Time / Endless Loop
Recording Intervals
1/2/5/10/15/20/30 seconds, 1/2/5/10/15/20/30/60 minutes
Battery
1 Lithium CR2 Battery
Battery Life
Maximum 8 months
Display
Temperature and Unit, Recording Settings, Battery Warning,
Amount of Data, Sensor Type and Recording Mode.
Data Capacity
8000 Readings x 2 Channels
Dimensions / Weight
H123mm x W58mm x D33mm / 132 grams
‑10 to 60℃/90%RH (No condensation)
Temp. and Humidity Resistance
Software ( ( ( ( ( Hi for Windows ) ) ) ) )
Channels Displayed
Up to 8 Channels simultaneously displayed / processed
Data Display
Graph, Table, High and Low Temp., Average Temp.,
Recording Interval, Amount of Data Recorded.
Data Output
Printed Graph, Printed Table, Text File
Control Functions
Programmed Start, Recording Interval, and Mode Settings.
Other Accessories
Communication Cable
1 RS-232C: D-SUB 9pin cable 1.5 meters
Battery
1 Lithium (CR2)
Attachment Plate
1 (includes 1 screw)
User's Manual
1 booklet
■ Inquires about the Product
Sasaga 5652-169 Matsumoto City NAGANO 399-0033 Japan
Tel. +81-263-27-2131
Homepage address : http://www.tandd.co.jp/english
E-mail :
overseas@po.tandd.co.jp
Fax. +81-263-26-4281
26