The Nova Max Pro Creat eGFR Control Solution should produce re-
sults that fall within the range of results printed on the vial label of
the Control being used. If the Control Solution test result is outside
the range (is either higher or lower), the Meter and Test Strip may
not be working properly as a system.

Important Information for Control Solution

• Use only Nova Max Pro Creat eGFR Control Solution.
• Check the expiration date on the Control Solution vial. Do not
use Control Solution past the expiration date, or you may get
inaccurate results.
• Use the Control Solution vial only for 90 days after the first opening.
When you open a new vial of Control Solution, count forward 90
days and write that date on the label of the Control Solution vial.
eGFR MeterIFU Manual PRO PN64322.indd 29
eGFR MeterIFU Manual PRO PN64322.indd 29
Running Nova Max Pro Creat eGFR Control Solution
29