Interfering Substances - NOVA BIOMEDICAL MAX PRO CREATININE eGFR Instructions For Use Manual

Table of Contents

Advertisement

A study was conducted to examine Creatinine and eGFR measurement
using the Nova Max Pro Creatinine and eGFR Meter System. The following
substances were tested and determined to have no clinical interference:
Tested Interfering Substances
Acetaminophen
Ascorbic acid
Bilirubin
Cholesterol
Creatinine
L-Dopa
Dopamine
Glucose
Heparin
Ibuprofen
D(+) Maltose Monohydrate
Methyl Dopa
Salicylate
Tolazamide
Tolbutamide
Triglyceride
Uric Acid
eGFR MeterIFU Manual PRO PN64322.indd 3
eGFR MeterIFU Manual PRO PN64322.indd 3

Interfering Substances

Tested Concentration Level
20.0 mg/dL
2 mg/dL
20 mg/dL
500 mg/dL
15 mg/dL
1 mg/dL
5 mg/dL
600 mg/dL
500 units/dL
48 mg/dL
500 mg/dL
2 mg/dL
48 mg/dL
25 mg/dL
25 mg/dL
750 mg/dL
20.0 mg/dL
3
1/14/22 3:42 PM
1/14/22 3:42 PM