Interfering.substances - NOVA BIOMEDICAL StatStrip Xpress Instructions For Use Manual

Lactate hospital meter
Table of Contents

Advertisement

Interfering Substances
Lactate Interferences:
The Nova StatStrip Xpress Lactate Hospital Meter exhibits
no interference from the following substances up to the
following concentration levels:
Substances
Acetaminophen
Ascorbic Acid
Bilirubin
Cholesterol
Creatinine
Dopamine
Ephedrine
Glucose
Ibuprofen
14
Introduction
Conc
mg/dL(mmol/L)
20.0
(0.66)
10.0
(0.57)
15.0
(0.26)
500.0 (12.9)
6.0
(0.53)
10.0
(0.53)
0.9
(0.055)
900.0 (49.7)
48.0
(2.33)
Substances
L-Dopa
Methyl-Dopa
Salicylate
Tetracycline
Tolazamide
Tolbutamide
Triglycerides
Uric Acid
Conc
mg/dL(mmol/L)
5.0
(0,25)
1.0
(0.042)
30.0
(1.87)
30.0
(0.62)
15.0
(0.48)
45.0
(1.67)
750.0 (8.78)
20.0
(1.05)

Hide quick links: