Chair's Preparaion - Promeba PS-251 User Manual

Chair, power track, locking system
Hide thumbs Also See for PS-251:
Table of Contents

Advertisement

USER GUIDE //
CHAIR PS-251 + POWER TRACK PA-260 + LOCKING SYSTEM PA-281
Review 04/20
03
INSTALATION
03.2
Chair's preparation
On receipt of the product it is advised:
·
Remove the packaging and distribute the material such that all components are visible.
·
Verify that all components / parts in the attached list are present.
The appliance must be checked before each use in order to detect any malfunction and / or damage
caused by transport and / or storage. In particular, check:
·
General functionality of the device.
·
Cleaning the device (remember that lack of cleanliness can cause the risk of cross-infection).
·
Absence of cuts, holes, rips in the structure, including belts.
·
Correct fixing of all nuts, bolts and nuts.
·
Correct fixing of the belts.
·
Correct attachment of the belts.
·
State of use (moving parts, wheels, belts, covers)
·
Integrity of belts and covers.
·
Integrity of the components.
·
Integrity of the handles.
·
Lubrication of moving parts.
·
State of use of the wheels and braking system.
·
Operation of springs.
·
The emergency vehicle is equipped with a fastening system dedicated to the Promeba
product.
·
There are safety belts for the immobilization of the patient and they are intact and operating.
·
The welds are intact, without cracks or breaks.
·
No tube or sheet metal present curves or cracks
If the above conditions are met, the device can be considered ready for use; Otherwise, you must im-
mediately remove the device from the service and contact the manufacturer.