COD. DESCRIPTION OF THE ERROR
1
- THE BATTERY VOLTAGE IS LESS THAN 3V
- IS DISCONNECTED OR DAMAGED
2
- THE BATTERY VOLTAGE IS SUPERIOR TO 34V
3
- THE STRETCHER DOES NOT RECEIVE TEN-
SION FROM AMBULANCE
4
- THE BATTERY VOLTAGE IS LESS THAN 18V
5
- THE TRACK RECEIVES TENSION THROUGH
THE TERMINAL CONNECTION TERMINALS
SUPPORT BUT NOT HAVE BEEN PRESSED
COMPLETELY
6
- BATTERY IS NOT DETECTED
7
- FAILURE OF ENGINE ENCODER
8
- INTENSIDAD SUPERIOR A 48V
9
- INTENSITY UPPER 48V
10
- FAILURE OF THE LOADING PROCESS
IMPORTANT CONSIDERATIONS
The electronic track system will automatically shut down if no operation has been performed in 15 minutes.
In up and down maneuvers repeatedly of up to 10 cycles uninterruptedly we must make a stop
Mandatory for at least 15 seconds for the motor-reducer system of the track to recover.
It is advisable to remove the battery from the interior of the track if it is going to go through long periods
of inoperative.
It is important to maintain a minimum of charge cycles every month to maintain good battery capacity.
Do not leave the battery uncharged for more than 2 MONTHS. It could stop working.
POSSIBLE CAUSE
- BATTERY DISCONNECTED
- BATTERY DAMAGED
- OVERLOAD DURING BATTERY CHARGING
- FAILURE TO CONNECT SUPPORT-TRACK
- AMBULANCE BATTERY DOWNLOADED OR
DAMAGED
- POWER SUPPLY CABLES CONNECTED BE-
TWEEN THE SUPPORT AND THE AMBULANCE
- ONE OF THE COPPER TERMINALS OF THE
TRACK HAS NOT CORRECTLY LEFT
- THE BATTERY IS DOWNLOADED
- TRACK HAS NOT ENTERED PROPERLY
- MICROPOWER FAILURE
- BATTERY DISCONNECTED
- CONNECTING THE DAMAGED BATTERY
- VERY LOW BATTERY LOAD
- CONTACT WITH TECHNICAL SERVICE
- TRACK WITH WEIGHT OVERLOAD
- LOW BATTERY
- THE TRACK DOES NOT UP OR LOWER SOFT-
LY
- TRACK WITH WEIGHT OVERLOAD
- LOW BATTERY
- THE TRACK DOES NOT UP OR DOWN WITH
SMOOTHNESS
- CONTACT WITH TECHNICAL SERVICE
POSSIBLE SOLUTIONS
- CONNECTING THE BATTERY CORRECTLY
- CHECK CONNECTION OF THE BATTERY CA-
BLES TO THE ELECTRONIC PLATE
- REPLACE BATTERY
- COOL THE BATTERY FOR 3/4 HOURS
- USING THE CHARGER PROVIDED BY PROMEBA
- REPLACE BATTERY FOR A NEW ONE
- CHECK CONNECTION TERMINALS BETWEEN
THE CATERPILLAR AND THE SUPPORT
- CHARGING BATTERY OR REPLACING FOR A
NEW ONE
- CHECK CONNECTION OF THE CABLES BE-
TWEEN
ORUGA-SUPPORT-AMBULANCE
- CHECK THAT THE CONNECTION TERMINALS
OF THE TRACK ENTER AND EXIT FROM YOUR
ACCOMMODATION CORRECTLY
- LOADING BATTERY TO THE MAXIMUM LEVEL
- MAKE SURE THAT THE TRACK HAS BEEN FIT-
TED INTO THE SUPPORT AND THAT THE CON-
NECTING TERMINALS ARE WELL CONNECTED
- CHECK MICROPOWER STATUS
- ADJUST THE BATTERY TO YOUR BASE
- REPLACE FOR NEW BATTERY
- CHARGING BATTERY WITH EXTERNAL CHAR-
GER
- CONTACT WITH TECHNICAL SERVICE
- DO NOT LOAD TRACK WITH MORE THAN
180KG
- CHARGING BATTERY WITH EXTERNAL CHAR-
GER
- LOOK FOR POSSIBLE HITS, INTERFERENCES OR
PARTS LOCKED IN THE TRACK
- DO NOT LOAD TRACK WITH MORE THAN
180KG
- CHARGING BATTERY WITH EXTERNAL CHAR-
GER
- LOOK FOR POSSIBLE HITS, INTERFERENCES OR
PARTS LOCKED IN THE TRACK
- CONTACT WITH TECHNICAL SERVICE