Set Gas Detectors On Test Bays; Start Tests - ION shawcity ARA DOCK4 User Manual

Hide thumbs Also See for shawcity ARA DOCK4:
Table of Contents

Advertisement

Set gas detectors on Test Bays

1) Gas detectors must be installed in the Test Bays to test.
2) Push the gas detectors into the Test Bays until it is firmly fixed. The ARA DOCK4 can test up
to 4 detectors simultaneously.
WARNING: If gas detectors are not inserted correctly, it may lead gas leak and/or failure of test.

Start Tests

There are two types of operations depending upon which button is pressed.
-
Bump Test (Blue): Confirms the detectors' response to the applied gas.
-
Calibration (Green): Calibrates the detectors' measurement accuracy against the applied
gas.
1) After gas detectors are installed, press Bump Test Button of Calibration Button to start
tests.
2) A bump test will take 10~30 seconds, and calibration will take approximately 90 seconds.
3) While the test is active, the LEDs will turn ORANGE.
4) When the test has completed, the LEDs will turn GREEN for pass or RED for fail.
Every test also performs the following maintenance operations:
-
Save each detector's event logs
-
Change each detector's configurations (if Configuration allowed)
-
Upgrade each detector's firmware (if required)
-
Buzzer test
-
LED test
-
Hibernate detector (if model supported and Configuration allowed)
WARNING: Do not remove gas detectors until the test completes.