3. Operation Procedure

1) Confirm that the probe is cleaned,
disinfected and if necessary
sterilized.
2) Confirm that the Magnetic sensor
attachment and the Spacer for
EZU-RV2S is disinfected or
sterilized, when use RVS
(Real-time Virtual Sonography).
3) Connect the probe to the
ultrasound diagnostic scanner,
operate the scanner, and adjust
the image, all according to the
instructions given in the
operation manual for the
ultrasound diagnostic scanner
with which the probe is used as
connected.
4) Confirm the direction of the
probe. The relationship between
the direction of the probe and
image is shown in Fig. 2. The
Right-left orientation mark on
the image indicates the
direction of the index mark of the
probe.
5) If a biopsy method is required, refer to its instruction manual.
6) Under sterile condition, protecting the probe by using covers is
recommended. Some Latex material may create allergic reaction.
Please use allergy free material covers.
7) After using the probe, perform the reprocessing procedure in
accordance with the procedure stated in "5. Reprocessing procedure
of the probe" every time immediately after completing the ultrasound
examination.
8) After using the Magnetic sensor attachment and the Spacer for
EZU-RV2S, perform the reprocessing procedure in accordance with the
procedure stated in "6. Reprocessing procedure of the Magnetic Sensor
Attachment and the Spacer for EZU-RV2S" every time immediately after
completing the ultrasound examination.
9) Store the Magnetic sensor attachment and the Spacer for EZU-RV2S
in the environment indicated in "7.Maintenance and Safety
inspection".
Fig. 2
Relationship between the
directions of the probe and the
Right-left orientation mark
-6-
Index mark
Right-left
orientation mark
Q1E-EP0428