3.

Operation Procedure

1)
Confirm that the probe, the Magnetic sensor attachment, and the
Spacer for EZU-RV2S are disinfected and if necessary sterilized.
2)
Attach the attachment to the probe according to the chapter 4.
3)
Connect the probe to the ultrasound diagnostic scanner, operate
the scanner, and adjust the image, all according to the
instructions given in the operation manual for the ultrasound
diagnostic scanner with which the probe is used as connected.
4)
Confirm the direction of the probe. The relationship between the
direction of the probe and image is shown in Fig.2.
The Right-left orientation mark on the image indicates the
direction of the index mark of the probe.
Fig.2
Relationship between the directions of the probe
and the Right-left orientation mark
Index mark
Right-left orientation mark
-5-
Q1E-EP0416