TEMPERATURE CALIBRATION

1. Turn on the power switch for either the hot air gun or the soldering
iron, depending on which one you are calibrating.
2. Remove the hot air gun or soldering iron from their holders - they
need to be at the temperature you have set and not in sleep mode.
Note: The temperature for the hot air gun will vary based on the
distance from the source.
3. If the independently measured temperature varies from the
temperature set on the unit, press and hold the "Up" and "Down"
buttons for 2 seconds for the respective device you are calibrating
until the display flashes. "CAL" will light up on the screen and the
temperature being calibrated will flash.
4. Enter the measured temperature into the flashing display by rotating
the airflow adjustment knob.
5. Once the flashing temperature shows the correct temperature,
press the Air Adjustment Knob to save the calibrated temperature
and exit the programming mode.