Figure 9 AS 100 connected to the DUT; position the AS 100 on the IC package if there isn't enough free space above
GND
3.2 Starting the AE XXX optical receiver
- Connect the AE XXX to the oscilloscope (BNC plug).
- The receiver output supplies a zero-delay voltage of 10 V independent of the measuring range
that is selected on the AS XXX. Thus, the vertical deflection on the oscilloscope should be
0.5-2 V/div.
- Connect the power supply to AE XXX (plug-in power supply unit) lower LED "ON" must be on.
Connect Sensor and receiver with the optical fibre upper LED "Sync" must be on. Loose the
clamped joint on the receiver, insert the optical fibre completely and slightly fasten the loose joint.
connectors and
Figure 10 AE 300
- The oscilloscope must show an output voltage and/or signal form that corresponds to that applied
to the sensor input.