Model F8159-1012A 1 CHANNEL STRAIN GAGE SENSOR SIGNAL CONDITIONER Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
SERIES 8159 SIGNAL CONDITIONER MANUAL DESCRIPTION The Series 8159 is a microprocessor-based digital indicator capable of interfacing directly to low level strain gage transducers. An internal, high gain, fully differential amplifier and a 4-½ digit analog-to-digital converter combine to accurately digitize the input signal. A 5V @ 60mA or 10V @ 120mA, short circuit protected, transducer excitation supply is also provided.
SERIES 8159 SIGNAL CONDITIONER MANUAL Figure 2 - Series 8159 Rear Panel View INSTALLATION The Series 8159 enclosure is designated for panel mounting in a 1/8 DIN cutout. The cutout dimensions are shown below. Figure 3 - Panel Mounting Cutout Dimensions To panel mount the 8159, perform the following steps: 1.
SERIES 8159 SIGNAL CONDITIONER MANUAL WIRING Reference Figures 4 and 5 for TRANSDUCER and INPUT/OUTPUT (I/O) wiring information. Power is applied with a 3-prong AC power cord. The instrument is protected by a 250V, 250mA fast acting 5mm fuse. The fuse holder is an integral part of the input power connector. A spare fuse is provided in the fuse holder.
Page 10
SERIES 8159 SIGNAL CONDITIONER MANUAL Figure 5 - Input/Output (I/O) Wiring...
Page 11
SERIES 8159 SIGNAL CONDITIONER MANUAL Figure 6 - Location of Jumpers and Potentiometers (Top View) Warning: Remove AC power to the 8159 meter before removing any enclosure panels or making any internal adjustments or changes. Switch 1 Switch 2 Switch 3 Switch Switch Bridge...
Page 12
SERIES 8159 SIGNAL CONDITIONER MANUAL Warning: Remove AC power to the 8159 meter before removing any enclosure panels or making any internal adjustments or changes. Housing RCAL Resistor Rear Panel Figure 9 - Location of RCAL Resistor (Top View) Warning: Remove AC power to the 8159 meter before removing any enclosure panels or making any internal adjustments or changes.
SERIES 8159 SIGNAL CONDITIONER MANUAL SWITCH DEFINITIONS Mode Selection Switch (S1) S1 is used to cycle through the various set points, Hysteresis, and calibration values and decimal point locations. The sequence is as follows. The designators, in parenthesis, identify text that will be displayed, momentarily, followed by the value.
SERIES 8159 SIGNAL CONDITIONER MANUAL Decrement Switch (S3) This switch serves two functions, depending on the present status of S1. If the instrument is in the Mode Selection sequence, this switch will decrement the flashing digit. If the instrument is in its operating mode, S3 becomes an AUTO CAL switch. CAUTION: Depressing this switch will cause the instrument to recalibrate its full-scale reading.
SERIES 8159 SIGNAL CONDITIONER MANUAL Set Point Low 2 (SPL2) SPL2 is independent and functions identically to SPL1. SPL2 has its own relay contact outputs and front panel LED (L2). Hysteresis High (HH) HH is the hysteresis value for SPH1 and SPH2. The hysteresis value determines the number of counts the displayed reading must fall below SPH1 and SPH2 values before deactivating their respective relays and LEDs.
Page 16
SERIES 8159 SIGNAL CONDITIONER MANUAL PROGRAM INSTRUCTIONS Four front panel switches (S1, S2, S3, S4) allow the user to program Set Point, Hysteresis, Calibration Number (CAL) and Decimal Points and monitor minimum, maximum or real time values. 1. Depress MODE (S1) switch once. The text SPH1 followed by the current value of Set Point High 1 (SPH1) will be displayed.
SERIES 8159 SIGNAL CONDITIONER MANUAL REMOTE TARE (ZERO) Remote zero will tare out any residual value, such as that caused by fixturing a load cell into a test stand. It is activated by momentarily connecting pin 14 to pin 5 on the 15-pin I/O connector REMOTE REBOOT This feature is used to clear the high or low set points if required.
Page 18
SERIES 8159 SIGNAL CONDITIONER MANUAL 8. With no load on the transducer, verify the instrument reading is low. 9. Depress the INCREMENT (S2) switch and hold until the TARE LED illuminates. The readout should be ZERO +/- 1 digit. 10. Apply the FULL SCALE load to the transducer. If a strain gage simulator is used, set it to the desired mV/V setting.
Page 19
SERIES 8159 SIGNAL CONDITIONER MANUAL 4. Set the CAL NUMBER to the full-scale value provided with the RCAL resistor. This is the shunt calibration valve indicated on the sensor’s calibration certificate. Example: If the RCAL resistor supplied with the instrument creates an output equivalent to 80% of a full-scale 500-pound load, set the CAL NUMBER to 400.00, 0400.0 or 00400.
SERIES 8159 SIGNAL CONDITIONER MANUAL OPTIONS RS232 SERIAL COMMUNICATION OUTPUT The RS232 transmission is of the displayed reading and is transmitted once every display update, if the DSR line is TRUE (HIGH). The data format is described below. 1. No parity bit 2.
SERIES 8159 SIGNAL CONDITIONER MANUAL Figure 12...
Page 22
Model Number Revision: A 1 CHANNEL STRAIN GAGE SENSOR SIGNAL CONDITIONER 8159-1012A ECN #: 37693 Performance ENGLISH OPTIONAL VERSIONS Linearity ± 0.015 % FS ± 0.015 % FS Optional versions have identical specifications and accessories as listed for the standard model Environmental except where noted below.