Summary of Contents for Rigol RSA5065N-OCXO

  • Page 1 User Guide Vector Network Analyzer Applicable to RSA5000-VNA/RSA3000-VNA Jun. 2020 RIGOL TECHNOLOGIES CO., LTD. 1.800.544.2843 www.calcert.com sales@calcert.com...
  • Page 2 Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. ⚫ RIGOL reserves the right to modify or change parts of or all the specifications ⚫ and pricing policies at the company’s sole decision. Information in this publication replaces all previously released materials.
  • Page 3: Safety Requirement

    RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please follow the instructions specified in this manual to use the instrument properly.
  • Page 4 Do Not Operate with Suspected Failures. If you suspect that any damage may occur to the instrument, have it inspected by RIGOL authorized personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
  • Page 5: Safety Notices And Symbols

    RIGOL Safety Notices and Symbols Safety Notices in this Manual: WARNING Indicates a potentially hazardous situation or practice which, if not avoided, will result in serious injury or death. CAUTION Indicates a potentially hazardous situation or practice which, if not avoided, could result in damage to the product or loss of important data.
  • Page 6: Care And Cleaning

    RIGOL Care and Cleaning Care Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning Clean the instrument regularly according to its operating conditions. Disconnect the instrument from all power sources.
  • Page 7: Rsa Series Overview

    RIGOL RSA Series Overview RSA series is a new generation of cost-efficient real-time spectrum analyzer with high performance. With superb performance specifications and the clear user interface, the RSA series allows you to operate it through various ways, such as pressing keys on the front panel, using the touch screen, connecting the mouse and the keyboard.
  • Page 8: Document Overview

    RIGOL Document Overview Topics in this manual: Chapter 1 Quick Start This chapter introduces the front/rear panel and user interface as well as announcements during first use of the analyzer. Chapter 2 Functions of the Front Panel This chapter gives detailed function descriptions of the front panel keys in VNA mode.
  • Page 9 RIGOL VIII VNA User Guide 1.800.544.2843 www.calcert.com sales@calcert.com...
  • Page 10: Table Of Contents

    Contents RIGOL Contents Guaranty and Declaration ................. I Safety Requirement ................II General Safety Summary ................II Safety Notices and Symbols ..............IV Care and Cleaning ..................V Environmental Considerations ..............V RSA Series Overview ................VI Document Overview ................VII Chapter 1 Quick Start .................
  • Page 11 Contents RIGOL Quick Save ..................2-49 Cont ..................... 2-49 Single ................... 2-49 System Function .................. 2-49 System ..................2-49 File....................2-57 Recall ................... 2-61 Save ..................... 2-63 Chapter 3 Appendix ................3-1 Appendix A: RSA5000 Accessories and Option List ........3-1 Appendix A: RSA3000 Accessories and Option List ........
  • Page 12: Chapter 1 Quick Start

    Chapter 1 Quick Start RIGOL Chapter 1 Quick Start This chapter gives you a quick review about the appearance and dimensions of the RSA series, its front and rear panel, user interface, as well as announcements during first use of the analyzer.
  • Page 13: User Interface

    Chapter 1 Quick Start RIGOL User Interface 5 6 7 11 12 Figure 1-1 User Interface (VNA) Table 1-1 User Interface Icons Name Description Measurement result Displays the current measurement results for the marker (when no marker exists, the measurement results display frequency/span value).
  • Page 14 Chapter 1 Quick Start RIGOL configure network parameters. : unlocks the front panel keys; : locks the front panel keys. : unlocks the touch screen; : locks the touch screen. : indicates that no USB storage device is inserted; : indicates that a USB storage device has been inserted.
  • Page 15 Chapter 1 Quick Start RIGOL be updated nor displayed. — The letter with strikethrough and in blue color indicates that the trace is updating but not displayed. It is useful in trace math operation. ⚫ The third line displays the enabling or disabling status of the math operation function of each trace.
  • Page 16: Mode Setting

    Chapter 1 Quick Start RIGOL Mode Setting Mode RSA provides five working modes: GPSA, RTSA, VSA (option), EMI (option), and VNA (option). Press Mode or to select the working mode. You can also tap the function keypad to select the desired mode.
  • Page 17: Mode Setup

    Chapter 1 Quick Start RIGOL Mode Setup The Mode Setup menu is used to set parameters global to different measurements under all working modes. These parameters are independent of the currently running measurement, and they are global to all the measurements.
  • Page 18: Install The Option

    RSA series provides various options (for option information, refer to "Appendix") to expand the function of the spectrum analyzer. If you need to purchase the option, please contact RIGOL sales representative. After you have purchased the option successfully, you will get the corresponding key. Then perform the following operations to install the option.
  • Page 19 Chapter 1 Quick Start RIGOL :SYSTem:LKEY RSA5000-PA@8AD12B8EBC5DF492D1D4289B7CBA5B6150BF6F5D752 D645C36D74530B05F39B49C461B23A50D6C94A34E06782AC4380070 B0D1A86BA84E02768391FFD70C2103 VNA User Guide 1.800.544.2843 www.calcert.com sales@calcert.com...
  • Page 20: Chapter 2 Functions Of The Front Panel

    Chapter 2 Functions of the Front Panel RIGOL Chapter 2 Functions of the Front Panel This chapter describes in detail the front-panel function keys of RSA series and their associated menu functions in VNA mode. Contents in this chapter: Basic Settings ◼...
  • Page 21: Basic Settings

    Chapter 2 Functions of the Front Panel RIGOL Basic Settings FREQ Sets the frequency parameters of the analyzer. Press FREQ on the front panel to enter the frequency setting menu. You can also tap the function keypad at the top of the screen to select FREQ.
  • Page 22 Chapter 2 Functions of the Front Panel RIGOL Start Freq Sets the start frequency of the current frequency channel. Remarks: When you modify the start frequency, the span and center frequency will be changed. Table 2-2 Start Frequency Parameter Remarks...
  • Page 23 Chapter 2 Functions of the Front Panel RIGOL Stop Freq Sets the stop frequency of the current frequency channel. Remarks: When you modify the stop frequency, the center frequency and start frequency will be modified automatically if the span remains to be unchanged.
  • Page 24 Chapter 2 Functions of the Front Panel RIGOL Remarks: In Auto mode, the CF step is 1/10 of the span. ⚫ In Manual mode, you can use the numeric keys to set the step size. ⚫ VNA User Guide 1.800.544.2843 www.calcert.com...
  • Page 25: Span

    Chapter 2 Functions of the Front Panel RIGOL SPAN Sets the frequency range of the current channel. Press SPAN to enter the span setting menu. You can also tap the function keypad at the top of the screen to select SPAN.
  • Page 26: Ampt

    Chapter 2 Functions of the Front Panel RIGOL AMPT Sets the amplitude parameters of the analyzer. Press AMPT on the front panel to enter the amplitude setting menu. You can also tap the function keypad at the top of the screen to select AMPT.
  • Page 27 Chapter 2 Functions of the Front Panel RIGOL Ref Position Sets the reference position. When it is set to 5, the reference position is located in the middle of the graticule; when set to 0, it is at the bottom of the graticule; when set to 10, it is located at the top of the graticule.
  • Page 28 Chapter 2 Functions of the Front Panel RIGOL Smith 0.03 to 20 Polar 0.03 to 20 Expand Phase 1a to 100G degree (°) Positive Phase 1a to 100G degree (°) Return Loss (DTF) 100f to 100G Log Mag(DTF) 100f to 100G...
  • Page 29: Sweep And Function Settings

    Chapter 2 Functions of the Front Panel RIGOL Sweep and Function Settings Sets the bandwidth of the analyzer. Press BW to enter the bandwidth setting menu. You can also tap the function keypad at the top of the screen to select BW.
  • Page 30: Sweep

    Chapter 2 Functions of the Front Panel RIGOL Sweep Press Sweep to enter the sweep setting menu. You can also tap the function keypad to select Sweep. The menu sets sweep-related parameters, such as sweep time, sweep points, and sweep time mode.
  • Page 31 Chapter 2 Functions of the Front Panel RIGOL Table 2-12 Sweep Time Parameter Remarks Default S11/S21: 1.05946 s DTF: 1.48760s Range 1 ms to 4,000 s Unit s, ms, us, ns, ps Knob Step sweep time/100, Min = 1 us...
  • Page 32: Trace

    Chapter 2 Functions of the Front Panel RIGOL Sets the parameter related to the tracking generator (TG). Output Enables or disables the tracking generator. In VNA mode, TG is enabled by default. When performing the calibration, TG is required to be enabled. After TG is disabled, the analyzer will stop sweeping.
  • Page 33 Chapter 2 Functions of the Front Panel RIGOL Trace4: purple ⚫ For details, refer to the notes about "trace indicator" in "User Interface". Select the corresponding trace to set the relevant parameters for the specified trace. By default, Trace1 is selected, and the trace type is "Clear Write".
  • Page 34 Chapter 2 Functions of the Front Panel RIGOL Trace Display Sets the display contents of the selected trace. By default, it is Data. Data ⚫ Memory ⚫ Data&Memory ⚫ ⚫ Remarks: You need to perform Data->Memory operation first, and then the memory ⚫...
  • Page 35 Chapter 2 Functions of the Front Panel RIGOL marker position. Log/Phase: shows the magnitude (in dB) and the phase at the current ⚫ marker position. Real/Imag: shows the real and imaginary components at the current ⚫ marker position. R + jX: shows the real and imaginary components of the impedance at the ⚫...
  • Page 36 Chapter 2 Functions of the Front Panel RIGOL 12. Log Mag(DTF) Time-domain log magnitude display format in DTF. 13. Lin Mag(DTF) Time-domain linear magnitude display format in DTF. 14. Return Loss(DTF) Return loss in DTF. Math Function After performing Data->Memory operation, you can perform the calculation between measurement data and memory trace.
  • Page 37 Chapter 2 Functions of the Front Panel RIGOL LRD Window ⚫ UDR Window ⚫ Four Window ⚫ 2-18 VNA User Guide 1.800.544.2843 www.calcert.com sales@calcert.com...
  • Page 38: Trigger

    Chapter 2 Functions of the Front Panel RIGOL Trigger Selects the trigger source and sets trigger-related parameters. Source Sets "Free Run", "External 1", and "External 2" to be the trigger source. Free Run The trigger conditions are met at any time, that is, the analyzer generates trigger signals continuously.
  • Page 39 Chapter 2 Functions of the Front Panel RIGOL input via the [TRIGGER IN/OUT] connector on the rear panel. When the signal meets the set trigger conditions, trigger signals are generated. Note: The input signal frequency on the external trigger interface should not be greater than 1 MHz.
  • Page 40 Chapter 2 Functions of the Front Panel RIGOL Table 2-16 Trigger Hold-off Time Parameter Remarks Default 100 ms Range 0 μs to 500 ms Unit s, ms, μs, ns, ps Knob Step trigger holdoff time/100, Min = 1 μs Left/Right Arrow Key Step Up/Down Arrow Key Step at 1-1.5-2-3-5-7.5 step...
  • Page 41 Chapter 2 Functions of the Front Panel RIGOL triggers on the rising edge … … trigger s ignal … … sweep measurement1 trigger delay aut o trig trigger delay trigger occurs trigger occurs … … sweep measu rement2 (turn off trigger delay)
  • Page 42: Measurement Settings

    Chapter 2 Functions of the Front Panel RIGOL Measurement Settings Meas VNA provides measurement for S11, S12, and DTF. Press Meas to enter the measurement setting menu. You can also tap the function keypad at the top of the screen to select Meas.
  • Page 43 Chapter 2 Functions of the Front Panel RIGOL Table 2-18 Average Count Parameter Remarks Default Range 1 to 10,000 Unit Knob Step Left/Right Arrow Key Step Up/Down Arrow Key Step Ref Level Sets the reference level. It shows the max. power/level value displayed in the current graticule.
  • Page 44 Chapter 2 Functions of the Front Panel RIGOL Table 2-19 Aperture Step Width Parameter Remarks Default Range 1 to (sweep points/10) Unit Knob Step Left/Right Arrow Key Step Up/Down Arrow Key Step at 1-2-5 step Align In 1-port calibration, the calibration data are measured by connecting the OPEN standard, SHORT standard, and LOAD standard to the required test port.
  • Page 45 Chapter 2 Functions of the Front Panel RIGOL Remarks: Before calibration, set the required sweep frequency range first and select the ⚫ specified calibration kit. The calibration is only valid for the S11 measurement item. When switching to ⚫ S21 or DTF, you need to recalibrate.
  • Page 46 Chapter 2 Functions of the Front Panel RIGOL Table 2-21 Port 1 Length (m) Parameter Remarks Default 0.00 m Range -3.00 Gm to 3.00 Gm Unit Meter (m, dm, cm) Knob Step Port 1 length/100 Left/Right Arrow Key Step Up/Down Arrow Key Step...
  • Page 47 Chapter 2 Functions of the Front Panel RIGOL Cal Kits Before performing calibration, select one of the calibration kits. If you select a customized kit, define it by yourself. Selected Kit The kits supported by VNA include: CK106A ⚫ CK106E ⚫...
  • Page 48 Chapter 2 Functions of the Front Panel RIGOL Users can set the customized calibration kits. The parameters for different standard kits are shown in the following table. Table 2-25 Parameters for Different Standard Kits Offset Parts Type Length Parameter Parameter Open 0.0000mm...
  • Page 49: S21 Meas Setup

    Chapter 2 Functions of the Front Panel RIGOL S21 Meas Setup Avg Number Specifies the number of counts (N) for Average, Max Hold, and Min Hold. It is valid when Trace Type is "Average". The greater the value of N, the smoother the trace is displayed.
  • Page 50 Chapter 2 Functions of the Front Panel RIGOL For details about aperture step width, refer to descriptions in "Aperture Step Width" for S11. Align In full 2-port calibration, the calibration data are measured by connecting "Through" to two ports. This calibration effectively eliminates the frequency response reflection tracking error, directivity error, and source match error from the test set in a reflection test using that port.
  • Page 51 Chapter 2 Functions of the Front Panel RIGOL The relationship between port delay and port length is correlative, as shown in the following formula: Port Length = Port Delay * Velocity of Light * Velocity Factor Table 2-30 Port 1 Delay...
  • Page 52 Chapter 2 Functions of the Front Panel RIGOL Port 1 Velocity Factor Sets the Velocity Factor of Port 1. Table 2-33 Velocity Factor Parameter Remarks Default Range 0.1 to 1 Unit Knob Step Left/Right Arrow Key Step Up/Down Arrow Key Step...
  • Page 53 Chapter 2 Functions of the Front Panel RIGOL Table 2-36 Port 2 Length (m) Parameter Remarks Default 0.00 m Range -3.00 Gm to 3 Gm Unit m, dm, cm Knob Step Port 2 length/100 Left/Right Arrow Key Step Up/Down Arrow Key Step...
  • Page 54 Chapter 2 Functions of the Front Panel RIGOL Cal Kits Kits For details about the calibration kits, refer to "Cal " in "S11 Meas Setup". VNA User Guide 2-35 1.800.544.2843 www.calcert.com sales@calcert.com...
  • Page 55: Dtf Meas Setup

    Chapter 2 Functions of the Front Panel RIGOL DTF Meas Setup Opens the DTF parameter setting menu of the function selected in the Meas menu. The measurement results of DTF is displayed in the upper and lower windows. The upper window displays the relationship between transmission loss and ⚫...
  • Page 56 Chapter 2 Functions of the Front Panel RIGOL Clear Clears the calibration data acquired for the current measurement. Port Extension For details about the port extension, refer to "Port Extension" in S11. Cal Kits For details about calibration kits, refer to "Cal Kits" in S11.
  • Page 57 Chapter 2 Functions of the Front Panel RIGOL Sweep Point = 1 + Stop Distance * (span/1.5e8/Velocity Factor) (rounding up the number) Velocity Factor Sets the Velocity Factor of cable under test relative to the velocity of light in the vacuum.
  • Page 58 Chapter 2 Functions of the Front Panel RIGOL Window Function Spectral Amplitude Frequency Leakage Accuracy Resolution Gaussian Moderate Good Moderate Flattop Good Excellent Poor Rectangular Poor Poor Excellent Hanning Good Moderate Good Hamming Good Moderate Good VNA User Guide 2-39 1.800.544.2843...
  • Page 59: Marker Setup

    Chapter 2 Functions of the Front Panel RIGOL Marker Setup Marker Marker is used for marking the point on the trace. Reads the amplitude, frequency, or distance of a certain point on the trace via marker. Remarks: In VNA mode, each trace window provides 8 markers, and only a single marker ⚫...
  • Page 60 Chapter 2 Functions of the Front Panel RIGOL Remarks: If an active marker exists currently, then activate a reference marker at the ⚫ current marker; otherwise activate both the reference marker and Delta marker at the same time at the center frequency.
  • Page 61 Chapter 2 Functions of the Front Panel RIGOL marker in the time-domain trace for DTF measurement. When the trace you selected is in the DTF time-domain trace window, can this menu be unavailable. Table 2-46 Marker Distance Parameter Remarks Default 2.31 m...
  • Page 62: Marker To

    Chapter 2 Functions of the Front Panel RIGOL Marker To In DTF, when you select the time-domain trace window, Mkr->CF, Mkr->Start, and Mkr->Stop menus are grayed out and disabled. Mkr->CF Sets the center frequency of the analyzer to the frequency of the current marker.
  • Page 63: Peak

    Chapter 2 Functions of the Front Panel RIGOL Peak The peak search function enables the marker to move to the specific signal peak point, and then in combination with the function of Delta marker, it can provide a powerful analysis capability.
  • Page 64: Input/Output

    50 Ω. To measure a 75 Ω device, you should use a 75 Ω to 50 Ω adapter (option) supplied by RIGOL to connect the analyzer with the system under test, and then set the input impedance to 75 Ω.
  • Page 65 Chapter 2 Functions of the Front Panel RIGOL Parameter VNA Parameter Name FREQ Center Freq (Stop Frequency + Start Frequency)/2 Start Freq 1 MHz Stop Freq Fmax CF Step Auto, Fspan/10 SPAN Span Fmax – 1 MHz AMPT Ref Value...
  • Page 66 Chapter 2 Functions of the Front Panel RIGOL Window Layout Four Window (S11); UD Window (S21/DTF) Mode Setup Global CF Off, 0 Hz Meas Measurement Function Meas Setup (S11) Average Count Reference Level 0.00 dBm Aperture Step Width Port Extension Port 1 Delay 0.00 ns...
  • Page 67 Chapter 2 Functions of the Front Panel RIGOL Offset Length 14.89 mm -2.71 fF 2.478 fF/GHz -1.97e-1 fF/GHz -2.09e-2 fF/GHz Meas Setup(DTF) Avg Number Ref Value 0.00 dBm Port Extension Port Delay 1.00 as Unit meter Port Length 0.00 cm...
  • Page 68: User

    Chapter 2 Functions of the Front Panel RIGOL LCD Switch LCD Backlight 100% Power Switch Default Beep Switch SCPI Display User Key Language English Not affected by Preset settings. Note User User-defined shortcut key. For some hard-to-find but commonly used function menu, you can define it as shortcut key (for the defining method, refer to the introduction in "User Key").
  • Page 69 Chapter 2 Functions of the Front Panel RIGOL Preset Functions include: select the instrument settings to be recalled after the analyzer is powered on ("Last" or "Preset"); set the preset type ("Default" or one of the items from "User1" to "User6"); save preset.
  • Page 70 Chapter 2 Functions of the Front Panel RIGOL Figure 2-2 LAN Parameter Setting The following items can be set by the front panel keys, the keyboard shortcut keys, the externally connected mouse, the touchscreen, or remote operation. 1) Mode Sets the mode of obtaining the IP address.
  • Page 71 Chapter 2 Functions of the Front Panel RIGOL The format of the subnet mask is nnn.nnn.nnn.nnn; wherein, the range of "nnn" is from 0 to 255. You are recommended to ask your network administrator for a subnet mask available. 4) Gateway Press Gateway and use the numeric keys to input the desired gate address.
  • Page 72 ⚫ 2. Option Info Displays the option information. RSA5000N provides many kinds of options to meet various measurement requirements. To purchase the corresponding option, please contact RIGOL. 3. Self Test Screen ⚫ Tests whether the screen has a dot defect by using six colors: blue, red, green, gray, white, and black.
  • Page 73 Chapter 2 Functions of the Front Panel RIGOL Single-point Touch ⚫ Enters the single-point touch test interface. Touch or use the mouse to make stretch or strike gestures on the screen, if the corresponding traces are displayed, then it indicates that the point on the touchscreen is valid. To exit the test, press Esc for three consecutive times.
  • Page 74 Chapter 2 Functions of the Front Panel RIGOL Executes the factory resetting operation. ⚫ Language RSA supports multi-language menu, Chinese and English built-in help and pop-up messages. Press this key to select the desired display language. Settings 1. Power Switch Sets whether the power switch on the front panel is effective or not.
  • Page 75 Chapter 2 Functions of the Front Panel RIGOL that have been read or not read. You can also use the touch screen or the externally connected mouse to tap or click the specified icon in the prompt message dialog box.
  • Page 76: File

    Chapter 2 Functions of the Front Panel RIGOL File RSA allows you to save various types of files to the internal or external memory, and recall them when necessary. Press File on the front panel to enter the file operation menu interface.
  • Page 77 Chapter 2 Functions of the Front Panel RIGOL Table 2-47 Descriptions of File Type File Type Format Suffix Name Limit LIMIT .csv/.lim Trace&State .bin/.dat/.sta/.trs Calibration .csv Data&State Screen Image IMAGE .jpg/bmp/png License LICENSE .lic Text TEXT .txt Note: RSA can only recognize files whose filenames consist of Chinese characters, English letters, or numbers.
  • Page 78 Chapter 2 Functions of the Front Panel RIGOL Quick Print If the printer has been successfully installed, press this key to quickly print the currently displayed screen image. Print Prints the current screen according to print parameter settings when the printer has been successfully installed and is in idle state.
  • Page 79 Chapter 2 Functions of the Front Panel RIGOL 3. Select Device Selects the desired printer. Import License In the file manager interface, select a file to be imported and then press Import License to import the file. System Update Press this key to update the analyzer software after selecting the update file in the USB storage device.
  • Page 80: Recall

    Chapter 2 Functions of the Front Panel RIGOL Recall RSA allows you to recall various types of files saved to the internal or external storage memory. Press Recall on the front panel to enter the file recalling menu. The available file types include: State, Trace&State, Calibration Data&State, and All (all of the above...
  • Page 81 Chapter 2 Functions of the Front Panel RIGOL 1. Recall Press File Explorer to enter the file manager interface. Select a file and then press Recall to recall the file. 2. To Trace Selects the file and recalls it to the selected trace.
  • Page 82: Save

    Chapter 2 Functions of the Front Panel RIGOL Save RSA allows you to save various types of files to the internal or external memory. Press Save on the front panel to enter the file saving menu. The available file types include: State, Trace&State, Calibration Data&State, Screen Image, and All Above...
  • Page 83 Chapter 2 Functions of the Front Panel RIGOL The default filename is determined by the following methods: If you haven't input a filename by pressing the Save As key, its default ⚫ filename is tracestate<n>.trs. Wherein, n refers to the file number of the filenames that are named based on the naming mode in the current file directory.
  • Page 84 Chapter 2 Functions of the Front Panel RIGOL You can also touch the screen or use the mouse to touch or click the Save As menu, and then input the filename with the keypad. All Above Press All Above to enter the file saving menu for all types of files. All the types of files (State, Trace&State, and Calibration Data&State) are saved to the file.
  • Page 85 Chapter 2 Functions of the Front Panel RIGOL If you have pressed the Save As key to input the filename abc, then the file ⚫ is saved to be "abc.jpg". After that, press the key again, another file is saved automatically with the filename "abc1.jpg".
  • Page 86: Chapter 3 Appendix

    Chapter 3 Appendix RIGOL Chapter 3 Appendix Appendix A: RSA5000 Accessories and Option List Description Order No. Real-time Spectrum Analyzer, 9 kHz to 3.2 GHz RSA5032 Real-time Spectrum Analyzer, 9 kHz to 6.5 GHz RSA5065 Real-time Spectrum Analyzer, 9 kHz to 3.2 GHz...
  • Page 87: Appendix A: Rsa3000 Accessories And Option List

    Chapter 3 Appendix RIGOL Rack Mount Kit RM6041 USB Cable CB-USBA-USBB-FF-150 Note: For more details about the accessories and the options, please contact RIGOL sales representatives or the local distributors. Appendix A: RSA3000 Accessories and Option List Description Order No.
  • Page 88 VSWR Bridge, 2 GHz to 8 GHz VB1080 Near-field Probe NFP-3 Rack Mount Kit RM6041 CB-USBA-USBB-FF USB Cable -150 Note: For more details about the accessories and the options, please contact RIGOL sales representatives or the local distributors. VNA User Guide 1.800.544.2843 www.calcert.com sales@calcert.com...
  • Page 89: Appendix B: Warranty

    If a product proves defective within the warranty period, RIGOL guarantees free replacement or repair for the defective product. To get repair service, please contact with your nearest RIGOL sales or service office. There is no other warranty, expressed or implied, except such as is expressly set forth herein or other applicable warranty card.
  • Page 90: Index

    Index RIGOL Index All Auto Scale ......2-9 Next Peak ......... 2-44 Auto ........2-51 Next Peak Left ......2-44 Auto Scale ........2-9 Next Peak Right ......2-44 Center Freq ........ 2-2 Output ........2-13 CF Step ........2-4 Power On .........