CAL_RPT?
Returns a listing of a calibration report. (The same report as is printed to the serial port
when you use the CAL_PR command, but with no page formatting.)
Parameters
ACTIVE (The shifts resulting from the last calibration)
STORED (The shifts resulting from the last stored calibration)
CONSTS (A list of all calibration constants)
Response
(IAD) The report followed by the EOFSTR.
CAL_SHIFT?
Calculates and returns the difference (shift) in the results of updated calibration constants
at a specific input, range, and frequency.
Parameters
1. The shift to display from the list:
ACTIVE (the shift from STORED to ACTIVE)
STORED (the shift from ACTIVE to STORED)
2. Input binding post (WBND, INPUT1, INPUT2, or SHUNT)
3. The value at which to show the shift. This parameter selects the range the same way
as in the RANGE command.
4. Frequency at which to show the shift. Use 0.0 for dc.
Response
(Float) The shift in PPM
CAL_SKIP
Advances to the next step of an interactive procedure without executing the step (like a
CAL_NEXT except the step isn't executed). This is the same as when you press the "Skip
Step" softkey during front panel controlled calibration.
Parameter
None
CAL_SLST?
Calculates and returns the shift (as in CAL_SHIFT) for the given range and input at +DC,
-DC, and each service-calibrated frequency point.
Parameters
1. The shift to display from the list:
ACTIVE (the shift from STORED to ACTIVE)
STORED (the shift from ACTIVE to STORED)
2. Input binding post (WBND, INPUT1, INPUT2, or SHUNT)
Alphabetical List of 5790A Device-Dependent Commands
Remote Commands