E) Hold Function; F) Recording Minimum And Maximum Values; G) Rel Function - VOLTCRAFT VC-320 Operating Instructions Manual

Ac mini current clam
Table of Contents

Advertisement

Available languages

Available languages

e) HOLD function

• With the HOLD function, the current measurement value is shown on the display and the
measurement process is put on hold.
• Enable the function by pressing the HOLD/
the function.
• <
> appears on the display when the function is enabled.
• You can continue measuring after you have disabled the function.

f) Recording minimum and maximum values

• With the minimum/maximum value function, minimum and maximum values can be
registered and displayed on the LCD display.
• Enable the function by pressing the MAX/MIN button (3). Disable the function by pressing
the MAX/MIN button for approximately 2 seconds. The values on the display disappear.
• Minimum value: To display minimum values, keep pressing the MAX/MIN button until
< MIN > appears on the LCD display.
• Maximum value: To display maximum values, keep pressing the MAX/MIN button until
< MAX > appears on the LCD display.
• Difference: If you wish to view the difference between the maximum and the minimum
values, keep pressing the MAX/MIN button until < MAXMIN > appears on the LCD display.

g) REL function

• The REL function allows you to take a benchmark measurement to avoid potential line
losses. The current display value is set to zero and saved as a new benchmark value.
• Press the REL button (6) to enable the function. < REL > appears on the display. Press
the REL button again to turn the function off. < REL > disappears from the display. The
benchmark value is reset when the function is turned off.
• The LCD display shows the measured deviations from the benchmark value.
button. Press the button again to disable to
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vc-320-d1307543

Table of Contents