Optional Test VI: E6007A Visual Fault Finder Submodule
Figure 89 Measurement of the Output power
Set up multimeter
E6000C Mini-OTDR User's Guide, E0302
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2 Apply a 9/125 µ m patchcord with HMS-10 connectors.
3 Set the Multimeter:
dBm/W
dBm
λ = 635 nm
wavelength
sampling time
T = 100 ms
Range
AUTO
4 Before you switch on the DUT, zero the Multimeter:
press {Z
} on the Multimeter.
ERO
D Single-Mode/Multimode Module