Test II. Event Deadzone
Figure 78 Event Deadzone Test Setup
Settings
E6000C Mini-OTDR User's Guide, E0302
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
D Single-Mode/Multimode Module Performance Tests
2 Turn on the OTDR, and after the self-test has passed,
recall the default settings and the resolution mode.
[Settings]<Recall..><Default Setting>
[Settings]<Optimize Mode><Resolution>
3 Set the linestyle to solid
{V
}<P
><D
IEW
REFERENCES
4 Make sure that the length unit is set to meters.
[C
]<L
U
ONFIG
ENGTH
NIT
5 Set the Start and Span to 0.00–10.00 km, and the
Averaging time to 3 min.
[Settings]<Range><0-10 km>
[Settings]<Avg Time><3 min> (see the note on
page 224).
[Settings]<Meas. Mode><Averaging>
6 Select the required wavelength.
[Settings]<Wavelength>
7 Either (Single-mode module)
Set the pulsewidth to 10 ns.
[Settings]<PulseWidth><10 ns>
T
>: OFF
OTTED
RACE
><M
[
]>
ETER
M
231