  • Page 1 MS-60 INSTRUCTION MANUAL Pyranometer ISO9060: 2018 Class B MS-60S ISO9060: 1990 First Class...
  • Page 2 A- 5. C om m un ic a t io n Spec if ic a t io ns [ M S - 6 0 S SD I - 12 ] A- 6. R ec al i br at i on EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 1...
  • Page 3: Important User Information

    For warranty terms and conditions, please contact EKO Instruments or your distributer for further details. EKO guarantees that the product delivered to customer has been tested to ensure the instrument meets its published specifications. The warranty included in the conditions of delivery is valid only if the product has been installed and used according to the instructions provided in this operating manual.
  • Page 4 2- 3. A bo ut Op er ati ng Ma nu al Copy Rights Reserved by EKO INSTRUMENTS CO., LTD. Making copies of whole or part of this document without permission from EKO is prohibited. This manual was issued: 2019/07/30 Version Number: 4 2- 4.
  • Page 5 EKO is a unique manufacturer who can offer calibration service for pyranometers and pyrheliometers in-house. Based on the applied calibration methods EKO provides the best quality solar sensor calibrations compliant to the international standards defined by ISO/IEC17025 / 9847 [Indoor method] and ISO9059 [Outdoor method] [Certification: L13-94-R2 / www.pjlabs.com]...
  • Page 6 Date: April 4 , 2019 Position of Authorized Signatory: General Manager of R & D Department Name of Authorized Signatory: Shuji Yoshida Signature of Authorized Signatory: EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 5...
  • Page 7: Safety Information

    3. Safety Information EKO Products are designed and manufactured with consideration for safety; however, please make sure to read and understand this instruction manual thoroughly to be able to operate the instrument safely in the correct manner. WARNING Attention to user; pay attention to the instructions given on the CAUTION instruction manual with this sign.
  • Page 8 ➢ When carrying the MS-60 with sunscreen attached, always hold the instrument from the bottom. Holding only the sunscreen part may lead to dropping the sensor as it comes off from the sunscreen.
  • Page 9 It is fully compliant with the “Spectrally flat Pyranometer of Class B” of ISO9060: 2018 norm. The MS-60 Pyranometer is sensitive within the spectral range from 285 to 3000nm. It can be applied within the most demanding applications for solar radiation measurements.
  • Page 10: Package Contents

    Humidity and temperature sensor to monitor the sensor temperature and condition of the drying agents (silica gel) inside the sensor body. 4- 2. Pac k ag e Co nte nts Check the package contents first; if any missing item or damage is noticed, please contact EKO immediately. Table4-1. Package Contents Contents MS-60 / 60S ○...
  • Page 11: Getting Started

    Besides the mechanical protection, the optical properties of the glass domes minimize the undesired influence of long wave radiation on the shortwave radiation: the domes of the EKO pyranometers are only transparent for radiation emitted by the sun. Hence they block the undesired infrared radiation emitted by the Earth’s atmosphere.
  • Page 12 Durable materials are used for the cable and connector, and pin terminals are attached at the end of the cable for easy connection with data logger terminal block. * If longer cables, round terminals or fork terminals are required, please contact EKO Instruments or your distributor. [Also see [7. Specification, 7-4. Accessories List] for optional items.] The connector should be protected by the self-fusing tape when the pyranometers are used in the areas with the risk of the salt damage.
  • Page 13 The sunscreen can be removed by loosening the knurling screw and sliding it towards the spirit level direction. *When carrying the MS-60 with sunscreen attached, always hold the instrument from the bottom. Holding only the sunscreen part may lead to dropping the sensor as it comes off from the sunscreen.
  • Page 14 [Installing at Tilted Position] After the MS-60 is adjusted to horizontal position in levelled surface, install it on tilted mounting position. *When installing the instrument, do not remove the levelling feet or fixed feet; if the levelling feet are removed, it may cause abnormal output values due to the thermal effects from the mounting part.
  • Page 15 Install a fuse [0.5A] in series between the DC power supply [+] and the MS-60S connection as shown below for over current protection. Power supply fuse[0.5A] DC5V to 36V 0 to 1V input Measurement Device Fig. 5-3B. How to connect MS-60S EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 14...
  • Page 16 Install a fuse [0.5A] in series between the DC power supply [+] and the MS-60S connection as shown below for over current protection. Power supply fuse[0.5A] DC12V Data collection device SDI-12 Fig. 5-3E. How to connect MS-60S EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 15...
  • Page 17 Also see 7-3. Output Cable for cable arrangements. DC-12V+ Slave SDI-12 Data Data collection No.n Ground Device Slave Slave Slave No. 3 No.1 No.2 Figure 5-5. Communication Connection with SDI-12 EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 16...
  • Page 18 *When selecting 0-1V output, a precision resistor is required separately. The output depends on the accuracy of the resistor. Figure 5-6. Connector pin number of MS-80/80S ach number corresponds to the number in Table 5-4. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 17...
  • Page 19 The maximum output voltage can be calculated by multiplying the maximum solar irradiance with the calibration factor [e.g. when the sensitivity of the MS-60 pyranometer is about 10μV/W・m 0.010mV/W・m , the maximum output voltage is about 1,400W/m times 0.010mV/W・m...
  • Page 20 Hence, sampling rates and data reduction methods can be defined right at the beginning of the data acquisition process. The response time that is given in the specifications of the EKO pyranometers states the amount of time, which is necessary to reach 95% of the final measurement value. It is also possible to define a 63.2% response [which is equal to 1-1/e].
  • Page 21: Maintenance And Troubleshooting

    6. Maintenance & Troubleshooting 6- 1. Mai nt en a nc e Using the EKO pyranometers, accurate results can be obtained if the glass dome and the condition of the instrument are maintained properly. Environmental conditions, such as for instruments mounted near highly frequented traffic lanes or airports, may have a deteriorating effect on the materials.
  • Page 22 Any change in tilt position after the For the S-series with Modbus output installation affect Data Weekly check the tilt position. measurements due to the cosine response of the sensor. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 21...
  • Page 23: Calibration Method

    6- 2. C alib ra ti on an d Me as u re me nt U nc e rtaint y It is recommended to recalibrate MS-60 pyranometer once every 2 years in order to verify the good quality of the solar radiation measurements. Below explains about the calibration methods of EKO pyranometers and the calibration uncertainty.
  • Page 24 6- 3. Tro ubl es hoo ti n g Read the following in case of any sensor trouble. If any questions should remain, please contact EKO for further technical support. Table 6-3. Troubleshooting Failure Action Make sure that the sensor cable is connected properly to the instrument.
  • Page 25: Pyranometer Specifications

    7- 1. Spec i fic a ti ons 1. Pyranometer Specifications The comparison table below, Table 7-1, shows typical values for the characteristic parameters of the EKO Pyranometers and the corresponding values of the ISO 9060: 2018 standard. The content of the characteristic item is partly changed from ISO 9060: 1990.
  • Page 26 Sensor setting can be changed by connecting the sensor to a PC (Use the optional RS-485 to USB converter cable and download the free configuration software from the WKO website When 4-20mA is 0-1600W・m , 0-1V : 0-1600W・m (default) EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 25...
  • Page 27 B. Body Height 91.5 mm C. Levelling Screw Height 16 mm Φ96 mm D. Width [including Sunscreen/Cover] E. Overall Height [approx.] 107.5 mm Figure 7-1. Outer Dimensions EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 26...
  • Page 28: O Ut Pu T C Ables

    PC with USB terminal * The standard cable length is 10m *The setting can be changed with the RS485 / USB conversion cable and dedicated software. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 27...
  • Page 29 Pyrheliometer is classified into 4 classes of AA, A, B and C based on the performance of each characteristic, and specifications of "spectrally flat radiometer" and "Fast response radiometer" are set as sub-categories. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 28...
  • Page 30 [ISO 9060: 2018 added] Error generated when converting the analog output of Pyranometer processing or Pyrheliometer into a digital signal with a signal converter etc. errors EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 29...
  • Page 31: Software Installation

    A- 3. Co nfi gu r ato r So ft wa re [MS -6 0S] With the dedicated software which can be downloaded from the EKO website and optional communication cable for converter setting, configurations can be changed. 1. Software Installation Download the latest version “SensConf.zip”...
  • Page 32 Fig. A-4. Query Success 3] Main window open. 4] When “Read Memory” is pressed, configurations are read from the MS-60S memory and displayed on the window. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 31...
  • Page 33 Manufacturing date information [cannot be changed] Product serial number [cannot be changed] Name Product name [cannot be changed] Calibration Date/ Product calibration date, sensitivity [cannot be changed] EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 32...
  • Page 34 Reads the configuration from the file. The file name is “XXX.xml” Load File Writes the current configuration to the file. The file name is “XXX.xml” Save File COM Setting Opens the communication setting window. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 33...
  • Page 35 With code, 16 bit integer value UINT32 Without code,32 bit integer value SINT32 With code,32 bit integer value IEEE754 32bit Single precision floating FLOAT32 point A string[ASCII Code] EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 34...
  • Page 36 Analog input setting UINT16 Analog output setting 109, 110 Calibration value of analog input setting FLOAT32 9 types in combination of amplifier type and range 127, 128 EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 35...
  • Page 37 Calibration history 0[Calibration date,YYYYMMDD UINT32 type] 186, 187 FLOAT32 Calibration history 0[Sensitivity [μV/W/m 188, 189 Calibration history 1[Calibration date,YYYYMMDD UINT32 type] 190, 191 FLOAT32 Calibration history 1[Sensitivity [μV/W/m EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 36...
  • Page 38 Ex.] S/N SXXXXXX05: Address = 5, S/N SXXXXXX00: Address = 100 [when “00”, the address is ”100”] Serial Communication Setup [Address Number: 104] Value: 6 [Default] Select with number [3] from the combination of baud rate and parity bit. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 37...
  • Page 39 Correction value for analog output power Load Resistor for 0-1V Output [Address Number: 133, 134] Value: 100 [Default] Precision resistor value [Ω] used when analog 0-1V output is selected. EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 38...
  • Page 40 [k1: Address 172/173, k2: Address 174/175, k3: Address 176/177, k4: Address 178/179] Default Value {k1 : 0, k2 : 1, k3 : 0, k4 : 0} Setting that does not use linearity correction EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 39...
  • Page 41 Conversion example: Data conversion of power supply voltage and reference voltage [addresses 60 to 61] Conversion result: Power supply voltage 5000 [mV], reference voltage 2048 [mV] Address No. MS-60S Output Data Conversion result 12050 2048 EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 40...
  • Page 42 Next, calculate the data before conversion according to the IEEE754 standard. Conversion result: 820.52 [W/m2] Address No. Output Data Pre-conversion 444D2147 data Conversion result 820.52 EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 41...
  • Page 43 The device will return the W/m value. * Example of response by "al!" Command: “a18EKOMS-60vvvxx…xx<CR><LF>” EKO: Manufacturer, MS-60: Model name, vvv: Version number[3 digits], xxxxxxx: Serial Number EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4 Pg. 42...
  • Page 44 A- 6. Rec alib r atio n When MS-60 is calibrated at an external calibration laboratory, in practice slight differences can be expected relative to the MS-60 manufacturer calibration scale and lab-scale. A difference in the calibration measurement results can be explained by the differences with respect to the method of calibration, reference sensors, sensor characteristics and measurement conditions.
  • Page 45 EKO Asia, Oceania 1-21-8 Hatagaya, Shibuya-ku, Tokyo 151-0072 Japan P. +81.3.3469.6713 F. +81.3.3469.6719 info@eko.co.jp www.eko.co.jp EKO North America 111 North Market Street, Suite 300, San Jose, CA 95113, USA P. +1-408-977-7751 F. +1-408-977-7741 info@eko-usa.com www.eko-usa.com EKO Europe, Middle East, Africa,...