Operating pressure range
S, ST, T:
VAuto:
CPAP:
Supplemental oxygen
Maximum flow:
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable water tub:
Standard water tub, air tubing:
General
The patient is an intended operator.

Humidifier performance

Mask Pressure
RH output % at
cm H
O (hPa)
63°F (17°C) ambient
2
temperature
Setting 4
3
85
4
85
10
85
20
85
25
85
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
3 to 25 cm H
O (3 to 25 hPa)
2
4 to 25 cm H
O (4 to 25 hPa)
2
4 to 20 cm H
O (4 to 20 hPa)
2
For VAuto device: 4 L/min (CPAP, S, VAuto)
For S device: 4L/min (CPAP, S)
For ST device: 15 L/min (CPAP, S, ST, T)
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Water tub
7. Device
8. Inlet filter
5 years
2.5 years
6 months
RH output % at
Nominal system output AH
72°F (22°C) ambient
temperature
Setting 8
Setting 4
100
6
100
6
100
6
90
6
90
6
1
, BTPS
2
Setting 8
>10
>10
>10
>10
>10
English