Maximum Common 42 V peak to earth 1 V p-p between T1 and T2 Mode Voltage Compliance CE (2014/30/EU) / RoHS2 (2011/65/EU) IEC 61000-4 2:2009, Class B 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Never use or store this product with batteries installed, or change batteries, in an environment where explosive or flammable vapors or dust suspensions may exist. For thermocouple thermometers suitable for use in explosive environments, see TEGAM’s 921A or 922A Intrinsically Safe Thermometers.
Page 9
Static discharge through a connected temperature probe may cause instrument damage. Use care to avoid static discharge when handling the instrument or connected probes. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Use the original packing materials if reshipment is necessary. If any dents, broken, or loose parts are seen, do not use the equipment. Notify TEGAM immediately. Check that all items are present. If any items are missing, notify TEGAM immediately.
(E, J, K, or T); Momentarily (do not hold) press to save your selection and move to the next parameter; 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 12
Please take a moment to familiarize yourself with this Operation Manual to learn about all the features and benefits of your new TEGAM Thermocouple Thermometer. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
The display is a large, easy to read, dual LCD display, with an LED backlight for clear viewing in low-light conditions. It simultaneously displays temperature measurements for 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 14
Channel 1 trend indicators Channel 1 minus indicator Channel 1 indicator T2 Probe Offset available on model 912B only. Models 912B only. Figure 3: LCD Display Description 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Setup Menu; Press to cycle through parameters as shown in Figure until the desired parameter is reached; To change the value of the current parameter, press 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Figure 6: View Modes and Statistics Press to change view modes. For each mode, the active measurement or statistic result is displayed on the second line of the display. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Key designators followed by (1.5s), e.g. (1.5s) , indicate that the key should be pressed and held for 1.5 seconds, then released to access the desired function. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Probe offset measurement using an ice bath or thermowell should only be performed by personnel trained and qualified in the use of such instruments and related metrology methods. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Setup Menu; (1.5s) Press four (4) times for 911B, five (5) times for 912B, to cycle to the Open Wire Detection Off/On parameter; “OWD OFF” is flashing near the bottom of the LCD display, and the current Open Wire Detection status is displayed on Line 2.
The nOrl function provides standard rounding, round down <0.5° and round up ≥ 0.5° 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 22
A decreasing temperature (cooling) -20.6° will display -20° until -21.0° is reached. This is for data on the instrument display only. Data transferred via the TEGAM Thermometer link to the TEGAM Cloud or via Bluetooth from the instrument will not have any rounding applied.
Channel 1 of the UUT. For two channel UUTs using the Split Thermocouple Cable, connect the miniature thermocouple connector of one leg to the UUT Channel 1 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
UUT on. Set the UUT parameters as shown in Figure 11 below. Refer to Section 3.3, Setup Menu as necessary for UUT parameter setup instructions. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 25
15. Allow the UUT displayed voltage to stabilize before proceeding. 16. Press UUT to automatically adjust the UUT voltage reading to the Applied Voltage, ±0.001 mV. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 26
Channel 2 input. If substituting another thermocouple type, use the appropriate Thermocouple Cable for the selected thermocouple type. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 27
UUT to normal operation. If substituting another thermocouple type, set the Thermocouple Calibrator as appropriate for the selected thermocouple type. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Shuts down Replace batteries (see Section 2.5, Batteries are low or unexpectedly or Battery Installation and depleted will not power on Replacement) 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
To erase all locally stored measurement data and reset accumulated statistics, press (1.5s) See Section 3.12, Clear Function for instructions. Instrument parameters will be retained. Refer to Section 3.3, Setup Menu to set instrument parameters as desired. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Once you have verified that the cause of the malfunction cannot be solved in the field and the need for repair and calibration service arises, contact TEGAM customer service to obtain an RMA (Returned Material Authorization) number. You can contact TEGAM customer service via the TEGAM website, www.tegam.com...
When is the problem most frequent?, has anything changed with the application since the last time the instrument was used?, etc. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
TEGAM, Inc. warrants the calibration of this product for a period of two (2) years from date of shipment. During this period, TEGAM, Inc. will recalibrate any product, which does not conform to the published accuracy specifications.
0.8 mm in diameter, may be used. Fluke 7526A meets the Thermocouple Calibrator specifications of Appendix A. All “Specification (2-Sigma)” column values rounded up to nearest hundredth. Appendix A: Required Equipment 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Appendices B. E XPANDED NSTRUMENT NCERTAINTIES Thermocouple Type E 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...