1 V p-p between T1 and T2 Mode Voltage Compliance RoHS2 IEC 61000-4 2:2009, Class B EN 55022:2010+A1:2015, MIL-PRF-28800F, Class 2 Class A; EN 61000-4 3:2006+A2:2010, 10 V/m (80 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Static discharge through a connected temperature probe may cause instrument damage. Use care to avoid static discharge when handling the instrument or connected probes. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
To install or replace batteries: Required Tools: Phillips Head Screwdriver 1. Identify the battery compartment located on the back of the instrument (see Figure below); 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
3. The instrument will immediately display a temperature measurement for the connected channels. However, to ensure 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 11
Please take a moment to familiarize yourself with this Operation Manual to learn about all the features and benefits of your new TEGAM Thermocouple Thermometer. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Channel 1 and Channel 2, current thermocouple type and temperature unit, trend indicators for both Channel 1 and Channel 2, and a battery voltage indicator. In 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 13
When viewing statistics, time elapsed since start of statistics collection Active statistic T2 Probe Offset available on model 912A only. Models 912A only. Figure 3: LCD Display Description 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
T2 Probe Offset ±0.1 ° increments below, until the desired parameter is T2 Probe Offset available on model 912A only. reached; Figure 5: Setup Menu Parameters and Values 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Figure 6: View Modes and Statistics Press to change view modes. For each mode, the active measurement or statistic result is displayed on the second line of the display. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 16
If the instrument records invalid measurement data during the statistics session such as an over-range, under-range, or open input value, ———— will be displayed for each affected statistic result. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Neither arrow is visible when the measurement is stable. For best accuracy, always allow the measurement to stabilize before evaluating or recording the measured temperature. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Probe offset measurement using an ice bath or thermowell should only be performed by personnel trained and qualified in the use of such instruments and related metrology methods. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 19
(912A only), or press to save the offset value and exit the Setup Menu. a. Alternatively, to disregard the new offset value and exit the Setup Menu, press 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
- UR - thermocouple type OPEn No probe is connected or the probe sensor is faulty ———— Cannot compute a valid statistical result Figure 9: Invalid Measurement Indications 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
150 to 760 °C ± 0.11 °C ± 0.16 °C 760 to 1200 °C ± 0.15 °C ± 0.22 °C Figure 10 continued on next page .. . 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 22
In the steps below, use the appropriate Thermocouple Cable and Thermocouple Calibrator settings as appropriate for the desired thermocouple type. 6. For two channel UUTs, both channels may be verified concurrently. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 23
16. Compare the Corrected Reading to the tolerances stated in the 2-Sigma Tolerance column of Appendix B to determine whether the UUT measurement is within published specifications. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
See Figure 13 for location. Figure 13: Alignment Access Hole Location 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 25
. The display will change to rES1 [2, 3 …] showing the actual measured value saved in the previous step. 18. Press UUT again. This will increment to the next CAL value. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 26
23. The UUT display will indicate as follows: a. Line 1: CALA b. Line 2: temperature in °C If substituting another thermocouple type, use the appropriate Thermocouple Cable for the selected thermocouple type. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 27
UUT to normal operation. If substituting another thermocouple type, set the Thermocouple Calibrator as appropriate for the selected thermocouple type. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Shuts down Replace batteries (see Section 2.5, unexpectedly Battery Installation and Batteries are low or depleted or will not Replacement) power on Figure 15: Common Troubleshooting Issues 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Once this information is prepared and sent with the instrument to our service department, we will do our part to make sure that you receive the best possible customer service and turnaround time possible. 4-10 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
When is the problem most frequent?, has anything changed with the application since the last time the instrument was used?, etc. 4-11 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
TEGAM service center with shipping charges prepaid. TEGAM Inc. shall pay for the return of the product to the customer if the shipment is to a location within the country in which the TEGAM service center is located.
International System of Units (SI) through the National Institute of Standards and Technology (NIST) or other recognized National Metrology Institutes. 4-13 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Appendices A. E XPANDED NSTRUMENT NCERTAINTIES Thermocouple Type E 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...