Sensor Type Configuration Temperature Unit Retention Offset Values Input Current ±50 nA 42 V peak to earth 1 V p-p between T1 and T2 Maximum Common 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
AVERTISSEMENT – RISQUE type et le modèle spécifique de batterie et D’EXPLOSION les pièces du fabricant mentionnées dans le manuel. L’utilisation d’autres batteries pourrait affecter la sécurité intrinsèque. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Check the integrity of the case prior to using the instrument in an Ex Hazardous • area. Look for any compromises like cracks or missing plastic. • Always comply with all safety codes. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 11
Do not attempt to measure temperatures beyond the range of the temperature probe. Probe damage or personal injury could result from exceeding a probe’s maximum temperature rating. Safety Notices and Information continued on next page ..10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
(This section, Use of Certified Thermocouple Probes, is controlled and may only be changed with approval of the Notified Body.) Use only TEGAM IS9 Series of thermocouple probes, with either the 921B or 922B, in potentially hazardous locations. Use of any other combination probe and cable, including thermocouple extension wiring, invalidates the type certification and may cause auto ignition in an explosive hazardous location.
Use the original packing materials if reshipment is necessary. If any dents, broken, or loose parts are seen, do not use the equipment. Notify TEGAM immediately. Check that all items are present. If any items are missing, notify TEGAM immediately.
UL 60079-11, Sixth Edition. Use of any other battery will invalidate the Type Certificate and create a hazardous condition. The only batteries that TEGAM, Inc. has secured such testing, which are approved for use in either a 921B or 922B, are listed on the Type Certificate. They are: •...
Page 15
Please take a moment to familiarize yourself with this Operation Manual to learn about all the features and benefits of your new TEGAM Thermocouple Thermometer. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
The display is a large, easy to read, dual LCD display, with an LED backlight for clear viewing in low-light conditions. It simultaneously displays temperature measurements for 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 17
Channel 1 trend indicators Channel 1 minus indicator Channel 1 indicator T2 Probe Offset available on model 922B only. Models 922B only. Figure 3: LCD Display Description 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Setup Menu; Press to cycle through parameters as shown in Figure until the desired parameter is reached; To change the value of the current parameter, press 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Figure 6: View Modes and Statistics Press to change view modes. For each mode, the active measurement or statistic result is displayed on the second line of the display. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Key designators followed by (1.5s), e.g. , indicate that the key should be pressed and held for 1.5 seconds, then released to access the desired function. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Probe offset measurement using an ice bath or thermowell should only be performed by personnel trained and qualified in the use of such instruments and related metrology methods. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Setup Menu; Press four (4) times for 921B, five (5) times for 922B, to cycle to the Open Wire Detection Off/On parameter; “OWD OFF” is flashing near the bottom of the LCD display, and the current Open Wire Detection status is displayed on Line 2.
The nOrl function provides standard rounding, round down <0.5° and round up ≥ 0.5° 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 25
For Example: An increasing temperature (heating) 20.6 ° will display 20 ° until 21.0 ° is reached. A decreasing temperature (cooling) -20.6 ° will display -20 ° until -21.0 ° is reached. 3-10 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Channel 1 of the UUT. For two channel UUTs using the Split Thermocouple Cable, connect the miniature thermocouple connector of one leg to the UUT Channel 1 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
UUT on. Set the UUT parameters as shown in Figure 11 below. Refer to Section 3.3, Setup Menu as necessary for UUT parameter setup instructions. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 28
15. Allow the UUT displayed voltage to stabilize before proceeding. 16. Press UUT to automatically adjust the UUT voltage reading to the Applied Voltage, ±0.001 mV. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 29
Channel 2 input. If substituting another thermocouple type, use the appropriate Thermocouple Cable for the selected thermocouple type. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 30
UUT to normal operation. If substituting another thermocouple type, set the Thermocouple Calibrator as appropriate for the selected thermocouple type. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Shuts down Replace batteries (see Section 2.7, Batteries are low or unexpectedly or Battery Installation and depleted will not power on Replacement) 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
To erase all locally stored measurement data and reset accumulated statistics, press (1.5s) See Section 3.12, Clear Function for instructions. Instrument parameters will be retained. Refer to Section 3.3, Setup Menu to set instrument parameters as desired. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Once you have verified that the cause of the malfunction cannot be solved in the field and the need for repair and calibration service arises, contact TEGAM customer service to obtain an RMA (Returned Material Authorization) number. You can contact TEGAM customer service via the TEGAM website, www.tegam.com...
When is the problem most frequent?, has anything changed with the application since the last time the instrument was used?, etc. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
TEGAM, Inc. warrants the calibration of this product for a period of two (2) years from date of shipment. During this period, TEGAM, Inc. will recalibrate any product, which does not conform to the published accuracy specifications.
Page 36
± 0.36 °C Thermocouple -200 to -101 °C ± 0.17 °C Type T -100 to -1 °C ± 0.12 °C 0 to 400 °C ± 0.11 °C 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
0.8 mm in diameter, may be used. Fluke 7526A meets the Thermocouple Calibrator specifications of Appendix A. All “Specification (2-Sigma)” column values rounded up to nearest hundredth. Appendix A: Required Equipment A-ii 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Appendices B. E XPANDED NSTRUMENT NCERTAINTIES Thermocouple Type B 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 39
Appendices Thermocouple Type E B-ii 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...