940A/945A IGITAL EMPERATURE ALIBRATOR HERMOMETER Model: 940A / 945A T/C Thermometer Operation Manual rev A Manual Part Number: 940A-900, Rev. A Published February 2019, Geneva, OH 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 2
FAR 12.211 (Technical data) and FAR 12.212 (Computer software) for the federal government, 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
-260°C – 400°C -120 to -70 ±0.4 -436°F – 752°F -70 to 80 ±0.3 80 to 400 ±0.2 See Graphs in Appendix B for Expanded Uncertainties 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 5
100 to 230 ±0.4 32°F – 4199°F 230 to 2315 ±0.3 Thermocouple types B, N, R, S, C, D, G, P, L and U are model 945A only. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 6
Keypad feedback Instrument retains: Sensor Type, Temperature Unit Power Cycle Configuration Offset Values, Presets, Statistics, Open Lead detection Retention status, 0%/100% span settings, and operating mode 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 7
3 AA Batteries, Quick Start Guide, Tilt Stand/Magnetic, Included Accessories Calibration Report HYSICAL HARACTERISTICS Dimensions 193 x 84 x 28 mm 7.6 x 3.3 x 1.1 in Weight including 362.9g 12.8 oz. Batteries 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Typical battery life under normal use conditions in laboratory environment. Continuous or repeated use of features such as the backlight or use or storage at high or low temperature extremes may reduce battery life. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Never use or store this product with batteries installed, or change batteries, in an environment where explosive or flammable vapors or dust suspensions may exist. For thermocouple thermometers suitable for use in explosive environments, see TEGAM’s 921A or 922A Intrinsically Safe Thermometers.
Page 11
Static discharge through a connected temperature probe may cause instrument damage. Use care to avoid static discharge when handling the instrument or connected probes. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Use the original packing materials if reshipment is necessary. If any dents, broken, or loose parts are seen, do not use the equipment. Notify TEGAM immediately. Check that all items are present. If any items are missing, notify TEGAM immediately.
The active thermocouple type is flashing on the display. Use to select the desired thermocouple type. You are setting the thermocouple type of the Read Channel.; 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 14
To save the current parameter value and exit the Setup Menu, press To disregard changes made to the current parameter value and exit the Setup Menu, press 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Figure 2 Keypad Button Functional Description , and keys have multiple functions which can be accessed by momentarily pressing the key, or alternatively, by pressing and holding the key for 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
TC input. Source output is matched to the reading using the same user settings. Bluetooth symbol on LCD reserved for future models. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 17
Under Range: The applied temperature is less than the minimum U rnG temperature for the selected thermocouple type Figure 4: LCD Error Indications 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
°C and °F”. If parameter 2 is set to “mV”, the remaining parameter choices available are in Figure 5 below under “Setup Choices for mV”. STATS are not active in mV mode. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 19
If on – Set Span Set 100% Level The 945A includes K,J,T,E,B,N,R,S,G,C,D,P,L,U All though still selectable, TC type is not visible on LCD while mV is the active unit 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Standard Deviation STDEV during the current session Standard Deviation is calculated using the population formula: = � ∑(−µ) Figure 6: View Modes and Statistics 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Auto-power off will remain disabled until instrument power is cycled. At next power on, auto-power off returns to the default enabled condition. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
SOURCE output = READ voltage + CJC voltage. This mode would be used for troubleshooting systems and readouts. The “Transfer” icon will blink and illuminate if selected. Figure 8: Operating Modes 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Current statistics will be invalidated after changing settings such as probe offset. Press (1.5s) to delete existing statistics data and initiate a new statistics session. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 24
Probe offset measurement using an ice bath or thermowell should only be performed by personnel trained and qualified in the use of such instruments and related metrology methods. 3-10 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
The LCD display will indicate CLEAr to confirm the action and return to active measurement mode. 3-11 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Once the desired preset is recalled, press the (1.5s) . The preset number should now be flashing. While the preset number is flashing, press (1.5s) . “CLEAr” will appear on the LCD 3-12 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Cannot compute a valid statistical result A thermocouple, shorted transducer, or other short circuit is Short ChAn1 plugged into the Source Channel during startup. Figure 10: Invalid Measurement Indications 3-13 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Suitable accuracy means a metrology-grade DMM. To achieve mV limits shown in Appendix C, the DMM must have accuracy equivalent to 30 ppm of reading and 9 ppm of range on the 100mV range. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 29
For “Source DC Volts Channel 1”, a PASS result is any “Measurement Result” value that is equal to or in between the Lower Limit and Upper Limit error numbers. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 30
An insulated box inside a calibration lab environment, see Appendix A. An accurate thermometer specification is 2 Sigma uncertainty ≤ .04°C (40mK) at the verification temperature. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 31
Customers performing CJC adjustments in their facility will need to calculate their own CJC uncertainty. Remove the UUT battery door housing to expose the alignment access hole. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 32
With the probe as close as reasonable to the channel 1 CJC, and temperature stabilized, enter the externally measured temperature in Celsius for channel 1. Repeat this step for Channel 2, CAL 12. See figure 13 above. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Unexpected statistics views until active reading on Statistics View Mode is active measurement is displayed (see Line 2 of Section 3.4 View Modes and Display Statistics) 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
(1.5s) . See Section 3.12, Clear Function for instructions. Instrument parameters will be retained. Refer to Section 3.3, Setup Menu to set instrument parameters as desired. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Once this information is prepared and sent with the instrument to our service department, we will do our part to make sure that you receive the best possible customer service and turnaround time possible. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
When is the problem most frequent?, has anything changed with the application since the last time the instrument was used?, etc. 4-10 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
TEGAM service center with shipping charges prepaid. TEGAM Inc. shall pay for the return of the product to the customer if the shipment is to a location within the country in which the TEGAM service center is located.
International System of Units (SI) through the National Institute of Standards and Technology (NIST) or other recognized National Metrology Institutes. 4-12 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 40
UUT. The opposite end shall be terminated with copper connections appropriate for the DMM. Straightened Required to access the calibration enable switch. Any rigid wire, approximately 0.8 Paper Clip mm in diameter, may be used. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Appendices B. E XPANDED NSTRUMENT NCERTAINTIES All uncertainty specifications for all charts are K = 2 unless otherwise noted. 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 42
Appendices Following graphs show total uncertainty in degrees C (k=2), with operating condition between 18-28 °C unless otherwise noted. B-ii 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 43
Appendices Thermocouple Type K B-iii 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 71
Published temperature uncertainty values can only be achieved if the temperature of UUT is known within 40mK. Customers performing CJC adjustments in their facility will need to calculate their own CJC uncertainty. Datasheet P/N 94X-DS Revision 1.0 C-ii 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...
Page 72
TEGAM INC. 10 TEGAM WAY GENEVA, OHIO 44041 CAGE Code: 49374 http://www.tegam.com WEB: C-iii 10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com...