3. Set ramp time
The ramp time controls the time of output
current that increase from 0 to the setting
level, and from the setting value back to
0. The ramp time is adjustable from 1 to 8
EN
seconds.
4. Set contraction time (ON-Time)
The ON time controls the contraction time.
By pressing the
control, the contraction
time is adjustable from 0 – 90 seconds. This
time includes the ramp time twice, once for
the ramp up and once for the ramp down.
Example: 10s ON time & 3s ramp time
= 3s ramp up, 4s full intensity, 3s ramp down
5. Set pause time (OFF-Time)
The O Time controls the time of relaxation
between two cycles. By pressing the
control, the relaxation time can be adjusted.
Both channels' stimulation is cycled on
and o by the contraction and relaxation
settings. The range is adjustable from 0 to
90 seconds. In Alternate mode, the OFF
time should be equal or more than the ON
time.
(OFF TIME ≥ ON TIME)
6. Set pulse width
The pulse width is adjustable from 30 to
400 uS. Press
control to enter this
setting. Then press the
or
control
to change the settings. If no instructions
regarding the pulse width are given in
therapy, set the control to 70 – 120 µs.