Page 2
AXION INTRODUCTION The STIM-PRO I-2000 is a medical device. The primary purpose is to stimulate muscles and nerves of the human body electrically through skin (transcutaneous) and mucous membranes (transmucosal) in order to strengthen the immune system and to alleviate incontinence problems.
Page 3
STIM-PRO I-2000 Urge incontinence (overactive bladder) - An overactive bladder is a sudden, unsuppressable urge to urinate, which forces the person affected to immediately go to a toilet. Mixed incontinence - This is a mixture of urgency and stress incontinence.
The device is a battery-operated pulse generator that gives electrical impulses to the body via electrodes and thus stimulates the nerves and muscles. The microprocessor of the STIM-PRO I-2000 generates electrical pulses whose pulse duration and frequency are preset. SAFETY INFORMATION Read the instructions carefully before using the device.
For hygiene reasons, the probe may only be used by one person. The unit must not be used if the pelvic floor is completely denervated. SCOPE OF DELIVERY Each STIM-PRO I-2000 is fully equipped with the following standard equipment: 1 piece EMS pelvic floor trainer 1 piece...
AXION STARTING THE STIMULATION The pelvic floor can be stimulated with either a rectal or vaginal probe, or with externally attached self-adhesive electrodes. Both applications lead to the desired effect. However, application with probes is more efficient and is therefore recommended.
Page 7
STIM-PRO I-2000 b) with self-adhesive TENS machine pads Connecting the cable Take the electrodes out of the packaging and clean the skin before applying to remove grease, sweat or cream residue. Connect the connecting cable to the device and to the electrodes.
AXION OPERATING THE DEVICE Turn on the device. Press and hold the On/Off button for a long time to turn the device on. During the application, you can pause and restart the application by briefly pressing this key. Program selection, start of stimulation In program mode you can choose between 5 different programs.
Page 9
STIM-PRO I-2000 If there is no noticeable muscle reaction, use more contact gel for probes or check the placement of the adhesive electrodes. If you have not received any prescription from your therapist or doctor, the recommended duration of treatment is 20 minutes at one session per day.
STIM-PRO I-2000 CHANGING BATTERIES To ensure the function of the device, please change the batteries when the low battery indicator is shown. Make sure the power is off. Open the battery compartment on the back of the device. Remove the batteries.
AXION MAINTENANCE, TRANSPORTATION AND STORAGE OF THE DEVICE Non-flammable cleaning solution (containing 70% of alcohol) is suitable for cleaning the device. Stains and spots can be removed with a cleaning agent. Do not submerge the device in liquids or expose it to large amounts of water.
There is no stimulation current? Check if the electrodes are connected to the same channel correctly and if the intensity is high enough. CONFORMITY TO SAFETY STANDARDS The STIM-PRO I-2000 device is in compliance with the EN 60601-1-2:2007 and 60601-1:2006 safety standards.
AXION TECHNICAL DESCRIPTION Channel 1 channel Intensity Adjustable 0 - 80 mA with a load of 500 Ohm on each channel Pulse amplitude Asymmetrical, bi-phasic square pulse Output Voltage 0 - 100 V Power Source 2 pieces 1.5V AA batteries Size 9.2cm(L) x 6.2cm(W) x 2.9cm(H)
STIM-PRO I-2000 WARRANTY All TENS models carry a warranty of 24 months from the date of delivery. The warranty applies to the stimulator only and covers both parts and labor relating thereto. The warranty does not apply to damage resulting from improper handling, the failure to follow the operating instructions, loss or dropping.