Non-Volatile Memory; Transducer Electronic Data Sheet (Teds) Interface; Filtering; Output Filtering - PCB Piezotronics ICP 483C30 Installation And Operating Manual

Signal conditioner
Table of Contents

Advertisement

MODELS 483C30, 483C50, 482M179 AND 483M217 GENERAL OPERATION MANUAL
2-9.

Non-Volatile Memory

This feature provides a mechanism to save and restore a programmed configuration. The programmable settings are
saved when the unit receives a SAVS command and restored when the unit powers up. The unit's non-volatile memory
may be reset to the factory default settings by using the RSET Command.

2-10. Transducer Electronic Data Sheet (TEDS) Interface:

TEDS sensors attach to the input connectors the same way a traditional ICP® sensor is attached to the signal
conditioner. The TEDS interface requires the Ethernet computer interface, and the input mode option. The digital
control and input mode features automatically perform the TEDS read or write function. After the communication with
the TEDS sensor is complete the input mode returns to the input mode (charge or ICP®) prior to the communication.
The current TEDS function will perform a read of the "raw" TEDS data and return the 32 bytes of TEDS sensor
EEPROM data and the 8 byte Application register contents, if it was programmed, and returns the data in an ASCII Hex
representation.

2-11. Filtering

Most Model 482C's can be ordered with optional Input and/or Output Filters which may be enabled and disabled for
each individual channel. The standard options are described below.

2-11.1 Output Filtering

The Model 482C can be ordered with low pass or high pass Output Filters installed in any or all channels for the
removal of unwanted noise. Output filters are located at the end of the signal path, after any gain has been applied. The
available Output Filter modules are fourth-order Butterworth low pass (-24 dB/octave) filters. High pass filters are
available upon request. The Output Filters are factory-installable only. Typical cutoff frequencies include:
The user may also define a specific cutoff frequency for the Output Filters that satisfies a specific need. Contact the
factory regarding individual requirements and model ordering code.
Output Filter
Filter Type
Option Number
00
Low Pass
01
Low Pass
02
Low Pass
03
Low Pass
04
Low Pass
05
Low Pass
06
Low Pass
07
Low Pass
08
Low Pass
09
Low Pass
10
Low Pass
11
Low Pass
12
Low Pass
13
Low Pass
14
Low Pass
Cutoff Frequency
(-3 dB)
No Filter Installed
100 Hz
200 Hz
500 Hz
1 kHz
2 kHz
5 kHz
10 kHz
20 kHz
50 kHz
100 kHz
300 Hz
20 Hz
33 Hz
1650 Hz
8