MODELS 483C30, 483C50, 482M179 AND 483M217 GENERAL OPERATION MANUAL
2-3.

Input Fault Detection

These models monitor two input fault conditions; "short" and "open," which indicate problems with sensor input and is
displayed through the front panel LEDs. Either case implies that the sensor is NOT functioning properly. An input is
shorted when it has a ground path for the sensor excitation and open when the sensor fails to draw the excitation.
Two voltage comparators consist of a window comparator that has two reference voltages (V
for "short" and "open." When the sensor's bias voltage (V
LED lights.
NOTE: Red LED implies input fault.
2-4.

Input Interface

The input signal conditioning for each channel provides a unity gain buffer with high-input impedance amplifiers. With
the AC coupled option it also eliminates the DC bias from the input signal with a 10-second time constant.
2-5.

Gain

The Programmable Gain Amplification (PGA) block consists of a decimal gain amplifier. See Figure 5.

Figure 3 Input Fault Window Comparator with LED Indicator

Figure 4 Input Amplifier Configuration

Figure 5 Selectable Gain Amplifier Configuration
) exceeds the comparator range, the front panel input fault
bias
5
) representing thresholds
ref