Figure 4.7 Example Of Securing The Stp Pump (When Securing The Inlet Port Flange With Claw Clamps); Figure 4.8 Example Of Securing The Stp Pump (When Installing The Damper In The Inlet Port Flange) - Seiko Seiki STPH803C Instruction Manual

Turbomolecular pump system
Table of Contents

Advertisement

STP-H803/H1303 Series Instruction Manual
Figure 4.7 Example of securing the STP pump
Figure 4.7 Example of securing the STP pump
(When securing the inlet port flange with claw clamps)
3) When installing the damper in the inlet port flange
Refer to Table 4.2 for rotational torque.
In case of using a damper, secure the base with all 8 screw-holes for legs
or all 8 attached legs. Follow "CAUTION" on page 4-9 about legs and
bolts for securing the base.
When the base cannot be secured because of the equipment design, install
the pump with a torque restraint like the one shown in Figure 4.8 (b).
Use a damper only at the vertically upright position.
DO NOT remove the bolts and nuts attached to reinforce the damper.
Vacuum Equipment
(a) When securing the base
Figure 4.8 Example of securing the STP pump
(When installing the damper in the inlet port flange)
Damper Clamps
Secure the base
Hole to prevent
from rotating
Leg
(b) When installing not to rotate
4-12
Vacuum Equipment
Claw Clamps
Secure the base
Vacuum Equipment

Advertisement