There are two methods of operating the STP pump: MANUAL and
REMOTE.
Select one which fits your vacuum equipment.
5.5
Manual Operation
Slide the "MANUAL/REMOTE" changeover switch on the rear panel to
"J\.1.ANUAL."
5.5.1 Powering ON
1) Switch ON the breaker on the rear panel (to prevent incorrect operation, a
metal fitting is attached to the breaker.
fitting and secure it).
2 ) Press the "POWER ON/OFF" switch (the switch is thoroughly locked down).
The magnetic bearing functions and the rotor levitates (POWER ON state).
The "POWER ON/OFF" switch built-in lamp lights.
5.5.2 Starting the STP Pump
1) Press the "MOTOR START" switch on the front panel after powering ON to
start and accelerate the STP pump.
The "ACCELERATION" lamp lights (ACCELERATION state).
2) When the STP pump attains the rated speed of rotations, the
"ACCELERATION" lamp goes out, and the "NORMAL OPERATION" lamp
lights (NORMAL OPERATION state).
5.5.3 Stopping the STP Pump
1) Press the "MOTOR STOP" switch on the front panel to stop the STP pump.
The STP pump starts to decelerate and stop.
The "NORMAL OPERATION" lamp or the "ACCELERATION" lamp goes
out, and the "BRAKE" lamp lights (BRAKE state)
(when the number of rotations is less than approx. 2000 rpm while the STP
p ump is accelerating, the "BRAKE" lamp does not light).
2) The "BRAKE" lamp goes out when the number of rotations slows down to
approx. 1000 rpm.
STP-H600/H1000 Series Instruction Manual
Loosen the screw, lift the :metal
5-3

Advertisement

loading

Summary of Contents for Seiko STP-H1000

  • Page 1 STP-H600/H1000 Series Instruction Manual There are two methods of operating the STP pump: MANUAL and REMOTE. Select one which fits your vacuum equipment. Manual Operation Slide the "MANUAL/REMOTE" changeover switch on the rear panel to "J\.1.ANUAL." 5.5.1 Powering ON 1) Switch ON the breaker on the rear panel (to prevent incorrect operation, a metal fitting is attached to the breaker.