General WK2017SD User Manual page 11

Ph, orp, conductivity, tds, salt and do meter with datalogging sd card
Table of Contents

Advertisement

Fig. 3. How to fill an oxygen probe's head with electrolyte
PROBE PLUG
PROBE HANDLE
To calibrate an oxygen probe, use the following procedure
while referring to Figures 1 and 4.
Fig. 4. An oxygen probe's main components
4-1 — Probe plug
4-2 — Probe handle
4-3 — Temperature-sensing metal
4-4 — Probe cap
4-5 — Probe head protective cover
1. Plug the probe plug into the DO probe socket
(callout 1-14 of Fig. 1) on the top of the meter.
2. Power on the instrument by pressing the POWER button
on the front panel.
3. Press the L button (which has the word "MODE" stenciled above it) until the display
shows the term "do".
4. Press the M button (which has the word "Function" stenciled above it) to switch the
unit shown on the display from "mg/L" to "%O2".
5. Wait at least five minutes until the display readings stabilize.
6. Use two fingers to press the "REC/ENTER" and HOLD/ESC buttons at the same time.
7. Release your fingers once this screen
appears (indicating that air typically has
an oxygen content of 20.9%.
8. Press the REC/ENTER button. The display will then count down from 30 seconds to 0.
9. When the countdown is complete, press the M button once to return to the normal
display of "mg/L".
PROBE-FILLING
ELECTROLYTE
Fig. 3
Fig. 4
11