AESCULAP
USA
Instructions for use/Technical description
ELAN 4 electro control unit GA800
Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)
®
Effective

Advertisement

Table of Contents
loading

Summary of Contents for Aesculap ELAN 4

  • Page 1 AESCULAP ® Effective Instructions for use/Technical description ELAN 4 electro control unit GA800 Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)
  • Page 2 Effective Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)
  • Page 3 Effective 20 21 GA808 GA810 Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)
  • Page 4 Effective GA822 GA849 GD412804 GA395SU GA861–GA864 GA804 GA865–GA869 GA805 GA860 GA806 GA836 GA808 GA831 GA837 GA810 GA844 GA832 GA833 GA824 Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)
  • Page 5 AESCULAP ® ELAN 4 electro control unit GA800 Legend Type BF applied part 1 ELAN 4 electro control unit GA800 2 Display with touch control panel Foot control 3 Coolant and irrigation pump 4 Flap Terminal for equipotential connector according to...
  • Page 6 ELAN 4 electro transverse Motor direction indicator for clockwise rotation pre- selected The display depends on whether the motor is acti- GA836 ELAN 4 electro micro sag- vated or disabled. ittal saw Motor direction indicator for counterclockwise rota- GA837 ELAN 4 electro micro...
  • Page 7 Effective Display/control elements in the system set- Symbol Designation tings menu Reset device to factory settings Symbol Designation Go back to menu structure Call up the system settings menu Call up the submenu Leave the system settings menu Move down list User profiles Move back up list Applied part settings...
  • Page 8: Table Of Contents

    Operating error ........18 Working with the ELAN 4 electro system ....
  • Page 9: About This Document

    Note Additional instructions and steps regarding the accessory components, in The clinical success of the use of the ELAN 4 electro motor system is depen- particular on their connection and reprocessing, can be found in the dent on the knowledge and experience of the surgeon. The surgeon must respective instructions for use or supplement for the components.
  • Page 10: Product Description

    Control unit Call up the system settings menu The ELAN 4 electro control unit 1 is designed for main voltage ranges of Rotational direction (here clockwise rotation) 100 V~ to 120 V~ and 220 V~ to 240 V~ from 50 Hz to 60 Hz. The mains voltage is converted into safety extra low voltage for supplying the micro motors in the applied parts.
  • Page 11: Applied Part Types

    4.3.5 Identification of control elements on system components The ELAN 4 electro control unit is authorized for operation in the operat- Control elements on the system components of the ELAN 4 electro motor ing room.
  • Page 12: Connecting The Power Supply

    Effective 6.1.2 Connecting the power supply 6.1.7 Connecting the ELAN 4 electro foot control GA808 to the control unit DANGER Note Risk of death by electric shock! The plug connection of the foot control has a yellow coded ring and a filled- ►...
  • Page 13: Connecting The Elan 4 Electro Motor Cable Ga804/Ga805/Ga806 Or Elan 4 Electro Drill Tool Ga844 To Control Unit

    Note part or with the applied part locked ("Off" position). All ELAN 4 electro motor cables (GA804/GA805/GA806) can be used in Otherwise the control unit will not recognize the applied part, and a mes- combination with all applied parts that do not have a fixed connection with sage will appear on the display (except for drill tool GA844).
  • Page 14: Releasing The Applied Part For Operation ("On" Position)

    Always carry out a function check prior to using the product. applied part. WARNING 6.1.14 Disconnecting the ELAN 4 electro foot control GA808 from Risk of injury and material damage due to inadvertent activation when the control unit repositioning/moving the foot control! ►...
  • Page 15: Activating The Motor/Pump Control Panel

    Effective 6.3.1 Activating the motor/pump control panel 6.3.3 Activating the coolant and irrigation pump/“Flush” function (continuous flushing) Note ► Activate the pump control panel, see Chapter 6.3.1. Motor settings of applied parts that are not connected to a cable cannot be ►...
  • Page 16: Enable Application Part Using Foot Control Ga808 Or Wireless Foot Control Ga810

    Effective 6.3.5 Enable application part using foot control GA808 or wireless System settings menu foot control GA810 Note Note The system settings menu can only be called up when no applied part is Applied parts connected to a hand control cable cannot be operated via the being operated.
  • Page 17: User Profiles

    Effective 6.4.1 User profiles 6.4.2 Motor settings User profiles can be created and edited in the user profiles menu. The motor settings menu displays the applied part types, see Fig. 10. Note When the control unit is restarted, the default profile is loaded automati- cally.
  • Page 18: Device Settings

    Effective ► Reset motor settings to default settings 6.4.4 Wireless foot c1ontrol The wireless foot control GA810 can be connected to the control unit or – Press button “Factory reset” a. disconnected from the control unit via the menu “radio footcontrol”. –...
  • Page 19: Reusable Products

    To ensure reliable operation, the product must be maintained at least once a year. Contact your national B. Braun/Aesculap representative if you need these services, see Chapter 10. Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)
  • Page 20: Troubleshooting List

    Effective Troubleshooting list  ► Have defective products repaired by Aesculap Technical Service, see  Chapter 10.   Display error messages  Faults that are recognized by the control unit are shown as error messages  in the display.
  • Page 21: Operating Error

    Effective 9.1.3 Operating error Message displayed Cause Remedy Motor turned off in operation. The motor was turned to the off-position n Stop the engine before turning off the engine The motor was turned to the off-position n opera- operation. tion. Stop the engine before turning off the engine Motor not recognized Applied part connected to control unit in "On"...
  • Page 22: Other Faults In Control Unit

    Motor coupling defective Change the applied part or motor cable. Tool cannot be coupled/decoupled ELAN 4 electro High speed: Auto- High speed tool doesn't engage Activate slider for tool release and matic tool lock blocked hold, then couple tool.
  • Page 23 Effective Malfunction Cause Detection Remedy Loud noise coming from applied part Loud, irregular noise Gearings/ball bearings defective Change applied part. Ball bearings of attachments Preventive measure: Ensure that the defective applied part is lubricated at regular intervals. High speed applied part vibrates Shaft of applied part bent Loud running noise / vibrations Change applied part.
  • Page 24: Fuse Replacement

    Effective 11. Accessories/Spare parts Fuse replacement DANGER 11.1 ELAN 4 electro motor cables, applied parts and Risk of death by electric shock! foot controls ► Unplug the device before changing the fuses! Art. no. Designation Specified fuse set: 2 units IEC 127 – T 6.3 A switching capacity H (1 500 A...
  • Page 25: Coolant And Irrigation Pump

    GA800 ELAN 4 electro control unit the coolant and irrigation pump GA804 ELAN 4 electro motor cable with hand lever Weight 9.5 kg ± 10 % GA805 ELAN 4 electro motor cable with hand control Dimensions (L x W x H) 380 mm x 330 mm x 201 mm ±...
  • Page 26: Ambient Conditions

    European Union, disposal is taken care of by the manufac- turer as a free-of-charge service. ► In case of questions regarding the disposal of the product, contact your local B. Braun/Aesculap representative, see Chapter 10. 14. Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc.
  • Page 27 Effective Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com ® AESCULAP – a B. Braun brand SOP-AIC-5001568 (TA014401) 2021-07 Rev. 3 Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)