PARAMOUNT BED cococia Riraku flow Instruction Manual

PARAMOUNT BED cococia Riraku flow Instruction Manual

Alternating pressure air flotation mattress

Advertisement

Quick Links

Alternating pressure air flotation mattress
cococia Riraku flow
Preface
Thank you for purchasing the Alternating pressure air flotation mattress cococia Riraku flow (hereinafter referred to as
the product).
z Before using the product, please read this instruction manual to fully understand the safe and proper operating procedures.
In addition, attendants and caregivers should also read this instruction manual.
z Store this instruction manual in a place that can be easily accessed for later reference.
z Users and/or carers are requested to report information about any serious incident related to this air mattress that occurs
within the EU to the manufacturer and the regulating authority of the member country of residence.
z 'Serious incident' means any incident that directly or indirectly leads, might have led or might lead to any of the following:
a) the death of a user, carer or other person.
b) the temporary or permanent serious deterioration of a user's, carer's or other person's state of health.
c) a serious public health threat.
z Some of the descriptions in this instruction manual may differ from the product you have purchased due to model improvements.
z Values described for the operating range, dimensions, angles, and weights are subject to normal manufacturing tolerances.
Except for special cases, words such as approximately and about are omitted in this instruction manual to improve readability.
z If you have any questions, please contact Paramount Bed or your distributor.
Intended Use
cococia Riraku flow is an air mattress with an internal pump, which is designed to disperse the body pressure to prevent pressure
ulcers in medical and elderly care facilities. Its operation panel, an accessory for cococia Riraku flow, is intended to operate the
alternating pressure air flotation mattress.
Note:
Patient target group: Persons weighing 20 to 180 kg who are at least 13 years of age
Persons requiring a function to disperse pressure applied to the body
Clinical benefit:
Body pressure dispersal to reduce the risk of pressure injury
Intended users:
Medical professionals and maintenance staff
* Consult with an expert, such as a doctor, nurse, or care counselor to prevent pressure ulcers. Pressure ulcers cannot be pre-
vented by the product alone. Full body control (such as nutritional status control or underlying disease control) is required.
* The product is one of the tools for general prevention of pressure ulcers. It is not intended to be used with a view to treatment or
alleviation of on-going pressure ulcers. It does not eliminate the need of posture change, either. Provide appropriate care such
as changing posture during the use of the product.
CONTENTS
1 SAFETY PRECAUTIONS  1
PROCEDURES 9
5 MAINTENANCE AND INSPECTION   18
6 TROUBLESHOOTING  21
8 REFERENCE INFORMATION   24
9 AFTER-SALES SERVICE  31
INSTRUCTION
MANUAL
7F13042000A1

Advertisement

loading
Need help?

Need help?

Do you have a question about the cococia Riraku flow and is the answer not in the manual?

Questions and answers

Summary of Contents for PARAMOUNT BED cococia Riraku flow

  • Page 1: Table Of Contents

    Intended Use cococia Riraku flow is an air mattress with an internal pump, which is designed to disperse the body pressure to prevent pressure ulcers in medical and elderly care facilities. Its operation panel, an accessory for cococia Riraku flow, is intended to operate the alternating pressure air flotation mattress.
  • Page 2 1 SAFETY PRECAUTIONS The safety precautions described in this instruction manual are intended to ensure safe use of the product and prevent personal injury and property damage. Precautions are classified into “WARN- ING” and “CAUTION,” which indicate the degree of hazard or injury that could result from improper use.
  • Page 3 WARNING z Do not use the product if the cover or padding is soiled z Inspect the product regularly (see page 20). or damaged. • For maintaining basic safety and essential performance in Excretion, blood, etc. could seep into the product, which may regards to EMC, the pump have an expected service life of cause infectious disease.
  • Page 4: Part Names  3

    * Part of ME EQUIPMENT that in normal use necessarily symbol definitions” on page 25 comes into physical contact with patient for ME EQUIPMENT * If the labels are removed or defaced, contact Paramount Bed to perform its function. (see page 31) or your distributor.
  • Page 5: Installation Procedure 4

    3 INSTALLATION PROCEDURE 1. Installing the mattress 1 When installing the mattress, make the bottom of the bed flat and check that there are no obstacles on the bottom of the bed. Quality label 2 The product has a top, bottom, head end, and foot end. When install- ing the mattress, make sure that the surface with the printed product Printed product logo logo faces up and that the quality label is at the foot end of the user.
  • Page 6 3 INSTALLATION PROCEDURE 2. Installing the pump and connecting a separately sold operation panel (accessory) z Install the pump properly according to the prescribed procedure, and check that the pump is con- nected and fixed to the mattress properly. Improper installation, connection, or fixing may cause the body pressure to not be dispersed effectively.
  • Page 7 The back raise/lower pressure relief function and bed power function can be used by connecting the product to a suitable bed specified by Paramount Bed using a bed link cable. When using the bed link cable, check how to connect to the bed and perform the wiring cor- WARNING rectly.
  • Page 8 3 INSTALLATION PROCEDURE ALiUS series Carefully read the instruction manual of the bed to learn how to operate the bed and ensure you understand how to safely and properly use the bed before operating it. 1 Make the bottom of the bed flat (horizontal). 2 Remove the power plug of the bed from the outlet and turn the battery off.
  • Page 9 A6 series Carefully read the instruction manual of the bed to learn how to operate the bed and ensure you understand how to safely and properly use the bed before operating it. 1 Make the bottom of the bed flat (horizontal) and raise the side-rails at the foot end.
  • Page 10: Functions And Operating

    3 INSTALLATION PROCEDURE 5. Making the bed 1 When the product has expanded, see the figure on the left and then cover the top and sides of the air mattress with a sheet. Towel, mattress pad, etc. * Use a towel, mattress pad, etc. as needed. 2 Insert the ends of the sheet into the sheet stoppers located at Sheet the four corners on the underside of the mattress.
  • Page 11 4 FUNCTIONS AND OPERATING PROCEDURES 2. Starting use „How to start Disconnect the power plug of the bed, then connect the bed link cable to the bed. See the procedure on pages 6 to 8. „How to shutdown Disconnect the power plug of the bed and turn the battery off, then disconnect the bed link cable from the bed. „Time required to complete air inflation When you insert the power plug into an outlet or connect bed link cable to bed, the air cells will be automatically inflated with air after a short while.
  • Page 12 4 FUNCTIONS AND OPERATING PROCEDURES 3-2. How to change settings „Automatic softness control setting Automatically change the internal pressure of the air cells to match the weight, body shape, and lying position of the user. The internal pressure of the air cells can also be changed manually. (The initial setting is automatic softness control, and the initial softness is 5.) [Switching between automatic softness control and manu- Status display screen (digital number) al softness control]...
  • Page 13 „Alternating pressure setting Repeatedly expand and contract sets of three air cells alternately and release the pressure on the body sequentially. (Initial setting: Enabled) „Operation sound setting Enable or disable the operation sound of the operation panel. (Initial setting: Enabled) „Dehumidify setting Eliminate any moisture and heat that have penetrated into the air mattress by periodically supplying air to the inside.
  • Page 14 4 FUNCTIONS AND OPERATING PROCEDURES 3-3. Max. inflate mode Increase the internal pressure of the air cells to make it easy to get on and off the bed, sit on the side of the bed, and change posture. In addition, the small flow air cell (SF cell) are contracted. When the user sits on the side of the bed, use the max.
  • Page 15 Using the BED Navi Use of the mattress may not be possible depending on the type of bed and BED Navi used. For details, contact Paramount Bed or your distributor. button Displays the home screen. 2 AIR MATTRESS button Displays the AIR MATTRESS home screen. (see page 15) * The figure shows an example.
  • Page 16 4 FUNCTIONS AND OPERATING PROCEDURES „ AIR MATTRESS home screen Touch the button in the home screen to display the AIR MATTRESS home screen. 1 Status display 5 SMALL FLOW SETTING button Displays information such as the air mattress operation sta- Changes the small flow setting.
  • Page 17 „ SMALL FLOW SETTING screen Touch the small flow setting button in the AIR MATTRESS home screen to display the SMALL FLOW SETTING screen if the small flow setting is ON, and the INFLATE SETTING screen if it is OFF. [SMALL FLOW SETTING screen (when the small flow setting is ON)] GO BACK (RETURN) button Confirms the settings and closes the SMALL FLOW SETTING...
  • Page 18 4 FUNCTIONS AND OPERATING PROCEDURES „ Function settings Touch the button in the AIR MATTRESS home screen to display the air mattress settings screen. 1 ALTERNATING PRESSURE button 6 AUTO MAX INFLATE SETTING button Sets alternating pressure. is displayed in the AIR MAT- Sets auto max.
  • Page 19: Maintenance And Inspection

    5 MAINTENANCE AND INSPECTION 1. Maintenance z Shutdown the air mattress by following procedure on page 10 before carrying out mainte- nance. There is a risk of malfunction of the product, possibly causing injury. z After maintenance, check that the glide sheet or air cells and covers (bottom cover and pump cover) are secured properly.
  • Page 20 Sodium hypochlorite „ Replacing the filter Regularly inspect and replace the filter. For the details on the dedicated filter, contact Paramount Bed (see page 31) or your distributor. 1 Press the hook of the filter case on the side of the pump and take out the case.
  • Page 21 To ensure safe use, inspect the product at least once a year. The product wears out and deteriorates differently depending on the frequency of use and environment. Regularly inspect the following items. When an irregularity has been noticed or when a detailed inspection is desired, contact Paramount Bed (see page 31) or your distributor. Part to inspect...
  • Page 22: Troubleshooting

    (see “How to shutdown” on page 10), and contact Paramount Bed (see page 31) or your distributor for repair.
  • Page 23 When not necessary, use the mattress with the max. inflate mode or automatic max. inflate mode turned off. Stop using the product and contact Paramount Bed or ― your distributor. • Is the back section of the bed raised? •...
  • Page 24: Long-Term Storage And Transportation

    “HXX.” A warning tone is output and the pump stops operating. (Major error indication only) * Stop using the product and contact Paramount Bed (see page 31) or your distributor. z The error code is also displayed on the BED Navi.
  • Page 25: Reference Information

    Power supply DC port: 24 VDC, 1.65 A / bed link port: 30 VDC, 1.32 A Service life 5 years (based on Paramount Bed data), excluding consumable items Essential performance: Adjust and maintain pressure in air cells Frequently used functions: Alternating pressure and changing position function „...
  • Page 26 8 REFERENCE INFORMATION 3. Product symbol definitions „Bed link cable Symbol Description Mattress Do not pinch or pull the bed link cable. „Quality label Symbol Description Bed link compatible Refer to the instruction manual. Refer to the instruction manual. Do not damage with a sharp object. Mark for restriction of the use of hazardous substances in electrical and electronic products in China...
  • Page 27 DS BED Navi/ BED Navi Operation panel (accessory) Symbol Description Symbol Description Max. inflate mode and auto max. inflate Press the button to display the air mattress mode settings screen of the BED Navi. Bed link function Automatic softness control Back support function Manual softness control Alternating pressure function...
  • Page 28 8 REFERENCE INFORMATION 4. Technical explanation regarding EMD (Electromagnetic Disturbance) The bed requires special safety precautions regarding EMD (Electromagnetic Disturbance). Install, operate, and use the bed according to the information provided below. The bed is intended for use in an electromagnetic environment specified according to “Electromagnetic disturbance.”...
  • Page 29 Guidance and manufacturer’s declaration ― electromagnetic emission The bed is intended for use in an electromagnetic environment specified below. The customer or the user of the bed must use this bed in an environment specified below. Emissions test Compliance Electromagnetic environment — guidance The bed uses RF energy only for its internal functions.
  • Page 30 8 REFERENCE INFORMATION Guidance and manufacturer’s declaration — electromagnetic immunity The bed is intended for use in an electromagnetic environment specified below. The customer or user of the bed must use this bed in an environment specified here. IEC 60601-1-2 Electromagnetic environ- Immunity test Compliance level...
  • Page 31 Guidance and manufacturer’s declaration — electromagnetic immunity The bed is supposed to be used in an electromagnetic environment where RF radio communication devices are used. The electromagnetic interference between the RF communication device and the bed can be prevented by maintaining the minimum distance between the two as recommended below in accordance with the maximum output and frequency of the communication device.
  • Page 32: After-Sales Service

    Your name, organization, address, and phone number 2. Minimum stock availability period Paramount Bed keeps consumable parts (necessary to maintain performance of the product) in stock for at least 6 years after the discontinuation of production. 3. Questions regarding after-sales service Contact Paramount Bed (see below).