Audible Alarm Settings and Service Functions
Menu Item 2 (Trend Clear)
Menu Item 3 (Language Selection)
Menu Item 4 (Baud Rate)
Menu Item 5 (EPP Mode)
34
When menu item 2 is selected, trend data that is available through the use of
menu item 1 will be deleted when the SatSeconds Alarm Limit button is
pressed and held until three beeps are heard.
At the present time only English is available.
Baud rates of 2400, 9600, or 19200 can be selected by first pressing the
SatSeconds Alarm Limit button when menu item 4 is displayed.
Use the Adjust Up or Adjust Down button to select the desired baud rate. The
baud rates will be displayed in the %SpO
or 192 (19,200). The protocol setting (menu item 5; EPP mode) determines
the baud rate that can be selected:
Option 1 ASCII = 19,200
Option 2 External equipment communications = 2,400, 9,600, 19,200
When the desired option is indicated in the %SpO
SatSeconds Alarm Limit button to initiate the current selection.
This menu item is used to change the method of sending data to the data port.
Two options can be accessed by first pressing the SatSeconds Alarm Limit
button when menu item 5 is displayed.
Use the Adjust Up or Adjust Down button to select the desired option.
Option 1 = real-time ASCII for printouts or displays
Option 2 = External equipment communications. Refer to the external
equipment manuals for the interfacing instructions.
When the desired option is indicated in the %SpO
SatSeconds Alarm Limit button to initiate the current selection.
window as 24 (2400), 96 (9,600),
2
display, press the
2
display, press the
2