AGFA XD 10 Series User Manual page 36

Detector calibration
Table of Contents

Advertisement

36 | DR Detector Calibration | Calibrating the DR Detector
Collima-
tion
On modalities that support automatic collimation, the collimation settings are
applied automatically and must be checked by the user.
Ralco collimators may limit the X-ray field. To overrule the limitation, the
collimator must be unlocked using a key and the collimation must be applied
manually.
Exposure Sequence
Procedure:
1. Set the mAs value to the indicative value for the exposure level.
2. Make exposures as instructed by the DR Calibration Tool.
0134I EN 20210924 1204
(collimation area should extend at least 1.5 cm on each side of
Note: If the exposure settings used for a previous calibration are
known, use these settings.
46 cm x 46 cm
the detector)