AGFA XD 10 Series User Manual page 32

Detector calibration
Table of Contents

Advertisement

32 | DR Detector Calibration | Calibrating the DR Detector
Collimation
On modalities that support automatic collimation, the collimation settings are
applied automatically and must be checked by the user.
Ralco collimators may limit the X-ray field. To overrule the limitation, the
collimator must be unlocked using a key and the collimation must be applied
manually.
Exposure Sequence
Procedure:
1. Set the mAs value to the indicative value for the first exposure level.
2. Make two exposures.
The DR Calibration Tool may instruct to increase or decrease the mAs
setting to reach the required exposure level.
3. Make 10 exposures.
4. Set the mAs value to the indicative value for the second exposure level.
5. Make two exposures.
The DR Calibration Tool may instruct to increase or decrease the mAs
setting to reach the required exposure level.
6. Make one exposure.
0134I EN 20210924 1204
The collimation area should extend at least 1.5
cm on each side of the DR Detector.
Note: If the exposure settings used for a previous calibration are
known, use these settings.