10 | DR Detector Calibration | Calibrating the DR Detector
DX-D 40C, DX-D 40G, DX-D
45C, DX-D 45G, DX-D 60C,
DX-D 60G
XD 10, XD 10+, XD 14, XD
14+, XD 17, XD 17+
DR 10e C, DR 14e C, DR 17e
C
DR 18M, DR 24M
Fixed DR Detector
0134I EN 20210924 1204
A single set of calibration data is used for all
NX workstations on which the DR Detector is
used. Choose one NX workstation that will be
used for calibration and perform the regular
calibration each time on the same NX work-
station.
Do not calibrate the DR Detector on another
NX workstation.
The DR Detector can only be used on a single
NX workstation.
Note: If the system has multiple DR Detectors, make a note of the ID
that is mentioned in the message on the NX workstation indicating
which of the DR Detectors must be calibrated. The ID is required to
identify the right detector during calibration.