Code
Description
Error
Blood glucose test strips have
been contaminated or used
Calibration code error
System error
Blood glucose test results
are higher than the detection
range
Solution
Replace it with a new one to re-measure.
Please re-insert the test strip and confirm
that the test strip matches with the machine.
Please detect with new glucose test strip;
if the problem persists, please contact with
dealer or manufacturer.
It indicates that you may have severe
hyperglycemia. Please immediately follow
the doctor's instructions for repeated
detection or treatment.
35
Code
Description
Error
Blood glucose test results
are lower than the detection
range
Ambient temperature is
above 40℃
Ambient temperature is
below 10℃
The battery is not enough and
the meter will turn off
automatically
Chapter 9: Other information
Solution
It indicates that you may have severe
hypoglycemia. Please immediately follow
the doctor's instructions for repeated
detection or treatment.
Please place the meter under a 10℃~40℃
environment for 30 minutes before detection.
Please place the meter under a 10℃~40℃
environment for 30 minutes before detection.
Please replace with new batteries
immediately