Chapter 7: Technical parameters and reference standards
Reference standards
●The result deviation between the blood glucose monitoring system and the
biochemical instrument (clinical lab) is within 95%, which shall follow below:
Precision:
Test range
≤4.2 mmol/L (≤75mg/dL)
>4.2mmol/L (>75mg/dL)
Accuracy:
Test range
<5.5mmol/L (<100mg/dL)
≥5.5mmol/L (≥100mg/dL)
Accuracy
Not more than ±0.83mmol/L (±15mg/dL)
Not more than ±20%
Precision
SD<0.42mmol/L (<7.7mg/dL)
CV<7.5%
29