Technical Data; Solvent Table (Excerpt); Warranty - IKA RV 8 Operating Instructions Manual

Hide thumbs Also See for RV 8:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 10

Technical data

Operating voltage range
Rated voltage
Frequency
Rated power of RV 8 device (without heating bath)
Rated power (max.) of switching adaptor
Speed
Speed tolerance
Set speed: < 100 rpm
Set speed: ≥ 100 rpm
Speed display
Dimensions of visible display area (W x H)
Display
Smooth start
Lift
Stroke
Head angle adjustable
Cooling surface
Cooling water flow rate min.
Cooling water flow rate max.
Cooling water pressure max.
Perm. On-time
Perm. ambient temperature
Perm. relative humidity
Protection acc. to DIN EN 60529
Protection class
Contamination level
Weight (no glassware; no heating bath)
Dimensions (W x D x H)
Operation at a terrestrial altitude
Subject to technical changes!
32
Vac
(100 – 240) ± 10%
Vac
100 – 240
Hz
50 / 60
W
75
W
90
rpm
5 – 300
rpm
± 1
%
± 1
mm
digital
37 x 18
7 segments display
yes
manual
mm
120
0° – 45°
cm
1500
2
l/h
30
l/h
100
bar
1
%
100
°C
5 – 40
%
80
IP 20
I
2
kg
12.5
mm
595 x 390 x 615
m
max. 2000

Solvent table (excerpt)

Solvent
Formula
Acetic acid
C
H
O
2
4
2
Acetone
C
H
O
3
6
Acetonitrile
C
H
N
2
3
N-Amylalcohol
C
H
O
5
12
n-Pentanol
C
H
O
5
10
n-Butanol
C
H
4
10
tert. Butanol
C
H
O
4
10
2-Methyl-2-Propanol
C
H
O
4
10
Butylacetate
C
H
O
6
12
2
Chlorobenzene
C
H
CI
6
5
Chloroform
CHCI
3
Cyclohexane
C
H
6
12
Dichloromethane
CH
CI
2
2
Methylenechloride
CH
Cl
2
2
Diethylether
C
H
O
4
10
1,2,-Dichloroethylene (trans)
C
H
CI
2
2
2
Diisopropylether
C
H
O
6
14
Dioxane
C
H
O
4
8
2
Dimethylformamide (DMF)
C
H
NO
3
7
Ethanol
C
H
O
2
6
Ethylacetate
C
H
O
4
8
2
Ethylmethylketone
C
H
O
4
8
Heptane
C
H
7
16
Hexane
C
H
6
14
Isopropylalcohol
C
H
O
3
8
Isoamylalcohol
C
H
O
5
12
3-Methyl-1-Butanol
C
H
O
5
12
Methanol
CH
O
4
Pentane
C
H
5
12
n-Propylalcohol
C
H
O
3
8
Pentachloroethane
C
HCI
2
5
1, 1, 2, 2, -Tetrachloroethane
C
H
CI
2
2
4
1, 1, 1, -Trichloroethane
C
H
CI
2
3
3
Tetrachloroethylene
C
CI
2
4
Tetrachloromethane
CCI
4
Tetrahydrofurane (THF)
C
H
O
4
8
Toluene
C
H
7
8
Trichloroethylene
C
HCI
2
3
Water
H
O
2
Xylene
C
H
8
10

Warranty

In accordance with IKA warranty conditions, the warranty period
is 24 months. For claims under the warranty please contact
your local dealer. You may also send the machine directly to our
factory, enclosing the delivery invoice and giving reasons for the
claim. You will be liable for freight costs.
Pressure for boiling point 40 °C in mbar
(For HB approx. 60 °C)
44
556
226
11
11
25
130
130
39
36
474
235
atm. press.
atm. Press.
atm. press.
751
375
107
11
175
240
243
120
335
137
14
14
337
atm. press.
67
13
35
300
53
271
357
77
183
72
25
The warranty does not cover worn out parts, nor does it apply
to faults resulting from improper use, insufficient care or
maintenance not carried out in accordance with the instructions
in this operating manual.
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rv 8 vRv 8 vcRv 8 flex

Table of Contents