Do not attempt to adjust the DDI or meter readings by turning the transducer zero or span
potentiometer (Refer to Page 4-1).
To prevent equipment damage, do not exceed the pressure rating of the transducer when
calibration is performed (Refer to Page 6-4).
Do not place any cleaning tools inside the pressure port connection of the sensor.
Damage to the diaphragm can result.
The pressure transducer is to be fully assembled prior to cleaning. (Refer to Page 6-6)
Do not immerse the pressure transducer or MS connector assembly in cleaning solutions.
Do not allow cleaning solutions to enter the pressure transducer (Refer to Page 6-6).