Page 4
3. Setup tripod or mast with CM20X Crossarm. Crossarm-To-Pole Bracket CM20X Crossarm...
Page 5
4. Mount gas analyzer and sonic anemometer. Leveling Bubble CSAT3A Sonic Anemometer Head EC155 Gas Analyzer Mounting Platform CM20X Crossarm EC155 Boom Adapter...
Page 6
5. Connect EC155 and CSAT3A to EC100 Electronics. EC155 Analyzer Cable EC155 Sample Cell Cable CSAT3A Cable...
Page 7
6. Connect system plumbing. Zero Air Tubing Pump Tubing Analyzer Tubing Span Gas Tubing Pump Module Cable...
Page 8
7. Connect system wiring and insert compact flash (CF) card. Insert CF Card System Power Cable (ensure +12 Vdc power supply is off) EC100 Power and SDM Cables...
Page 9
Power Cable Power Cable to +12Vdc to EC100 Power Supply (off) EC100 SDM SDM Cable to EC100 EC100 Cable Power Cable...
Page 10
Turn on the +12 Vdc power supply and use the datalogger keyboard display to set settings and initiate zero/span checks. (Follow the steps from left to right, top to bottom). 8.1 Press <Esc> to activate the display. Press 8.2 Select "Configure System." <Enter>...
Page 11
8.3 Select "SONIC_AZIMU." 8.4 Enter the CSAT3A sonic head azimuth.
Page 12
8.5 If the CPEC does not have a valve module, skip to the 8.10. Otherwise, select "Zero/Span Config." 8.6 Select "CO2_SPAN.PP: 0.00000."...
Page 13
8.7 Set CO2_SPAN_PPM to CO2 tank 8.8 Select CHECK_ZERO, SET_ZERO, concentration. CHECK_SPAN1, SET_SPAN1 in sequence.
Page 14
8.9 For each value selected on the previous screen, change the setting from "False" to "True" for auto zero/span. 8.10 Press <ESC> to exit.