Electrolytic Capacitor Measurement|For High Capacitance
DUT
Requirements
Measurement Setup
For high capacitance measurement, equivalent series
1
parameter Cs-D is commonly used.
Even though LVL:1 Vrms is set, less than 0.5Vrms
2
which is maximum appliable volatge for testing
electrolytic capacitors in accordance to JIS standard,
will be applied to the Aluminum electrolytic capacitor
under test.
Measurement Procedure
3-2