Reprocessing Procedure - Hitachi EUP-L65 Instruction Manual

Linear array probe
Table of Contents

Advertisement

5. Reprocessing Procedure

The probe, magnetic sensor attachment and spacer for EZU-RV2S must
be reprocessed after each use. Refer to the reprocessing instruction
in this chapter.
For reprocessing of the needle guide bracket EZU-PA6L1, the acoustic
coupler EZU-TECPL1 and the acoustic coupler attachment EZU-TEATC1
please refer to the corresponding instruction manuals.
WARNINGS
Limitations on
reprocessing
Transportation
before using
The probe is delivered unsterile. Prior to the
first use, reprocess the probe.
Temperature should not exceed 40°C during
reprocessing.
Probe connector is not water resistant.
The probe is not completely submersible. The
immersible part is shown in Fig.1. The un-immersible
part should be disinfected by wipe disinfection.
The probe should be packed in a sterile pouch or
container to transport from Central Sterile Supply
Department (CSSD) to an operating room. Be careful
not to damage the sterile pouch or container during
transportation.
- 14 -