Hitachi EUP-L65 Instruction Manual page 26

Linear array probe
Table of Contents

Advertisement

B)
Magnetic sensor attachment and Spacer for EZU-RV2S
1) Prepare the disinfectant solution as stated in the procedure for
the probe.
2) Immerse the magnetic sensor attachment and the spacer for EZU-RV2S
into the disinfectant (see Fig. 15). Set a clock to insure the
recommended contact time which is 5 minutes.
3) Rinse the magnetic sensor attachment and the spacer for EZU-RV2S
with deionized water for 1 minute. (alternatively: immerse them
in a tray filled with deionized water (see Fig. 15) for 5 min.)
4) Visually check the outer surface of the magnetic sensor attachment
and the spacer for EZU-RV2S for leavings of the disinfectant. If
necessary, repeat the rinsing.
Water
Detergent
Disinfectant
Fig. 15 Immersion of the probe, magnetic sensor attachment
and the elasto coupler
Do not immerse above
this line
- 20 -
Q1E-EP0739