Reprocessing Procedure

7.2
After each use of the probe, the reprocessing procedure including
cleaning, disinfection, and sterilization must be performed
immediately. A flow diagram showing the steps of the reprocessing
procedure is given in Fig.2. The reprocessing procedure must
be performed even after the probe was used with a probe sheath.
Cleaning Process
Cleaning
Rinsing and Drying
after cleaning
Liquid chemical
sterilization
/disinfection Process
sterilization/
disinfection
Rinsing and Drying
after sterilization/
disinfection
Do not sterilize the probe by Autoclave.
Autoclave sterilization causes the probe to malfunction.
DM 120918 v2
Users must choose either disinfection or
sterilization.
EtO sterilization
Fig.2 Reprocessing Procedure
WARNING
- 11 -
Process
Plasma sterilization
Process
Q1E-EP1441